FDA Recall Open, Classified

Vortex Disposable Forceps and Cannula 1. 23GA ACTU8 Forceps Adaptive VS0740.23 2. 25GA ACTU8 Forceps Adaptive VS0740.25 3. 25GA ACTU8 Forceps ILM VS0744.25 4. 25GA ACTU8 Forceps ILM -Myopic VS0744.25-35 5. 25GA ACTU8 Forceps Pro-Grip VS0713.25 6. 25GA ACTU8 Talon Forceps ILM VS0752.25 7. 25GA ACTU8 Talon Forceps Pro-Grip VS0751.25 8. 27GA ACTU8 Forceps ILM VS0744.27 9. 23GA Soft Tip Cannula VS0200.23 10. 23GA Soft Tip Cannula - 0.5mm VS0205.23 11. 25GA Soft Tip Cannula VS0200.25 12. 25GA Soft Tip Cannula - 0.5mm VS0205.25 13. 25GA Soft Tip Cannula - 0.5mm - Bulk VS0205.25B 14. 25GA Soft Tip Cannula-Bulk VS0200.25B 15. 27GA Soft Tip Cannula VS0200.27

Recall: Z-1247-2026 · Initiated December 16, 2025

Recall

Recall Number
Z-1247-2026
Event Number
98255
Firm
Vortex Surgical Inc.
FEI Number
3013002167
Product Code
HNR
Status
Open, Classified
Root Cause
Employee error
Initiated
December 16, 2025
Posted
February 4, 2026
Address
4 Research Park Dr, Ste 124, Saint Charles, MO, 63304-5639

Description

Vortex Disposable Forceps and Cannula 1. 23GA ACTU8 Forceps Adaptive VS0740.23 2. 25GA ACTU8 Forceps Adaptive VS0740.25 3. 25GA ACTU8 Forceps ILM VS0744.25 4. 25GA ACTU8 Forceps ILM -Myopic VS0744.25-35 5. 25GA ACTU8 Forceps Pro-Grip VS0713.25 6. 25GA ACTU8 Talon Forceps ILM VS0752.25 7. 25GA ACTU8 Talon Forceps Pro-Grip VS0751.25 8. 27GA ACTU8 Forceps ILM VS0744.27 9. 23GA Soft Tip Cannula VS0200.23 10. 23GA Soft Tip Cannula - 0.5mm VS0205.23 11. 25GA Soft Tip Cannula VS0200.25 12. 25GA Soft Tip Cannula - 0.5mm VS0205.25 13. 25GA Soft Tip Cannula - 0.5mm - Bulk VS0205.25B 14. 25GA Soft Tip Cannula-Bulk VS0200.25B 15. 27GA Soft Tip Cannula VS0200.27

Reason

There may be voids located in the seal of Tyvek pouches associated with identified catalog numbers and lots. A compromised sterile barrier has potential of bioburden contamination which could lead to infection.

Action

On December 16, 2025, the firm began notifying customers via "URGENT: MEDICAL DEVICE RECALL" letters. Customers were instructed to discontinue use of affected product and to return product for Vortex Surgical to evaluate the seals; use UPS account 23623Y to ship unused products to Vortex Surgical Inc for replacement or refund at: Attn: Andrew Ritts Recall Tyvek Seal, Vortex Surgical Inc., 4 Research Park Dr., Suite 124, St. Charles, MO 63304; complete and return the Acknowledgement and Receipt Form via mail or email to: [email protected]. Actions have been implemented to improve training of individuals associated with sealing and inspecting seals to prevent reoccurrence. The sealer associated with the issue is being updated to include a guard to prevent reoccurrence. If you have any questions, contact the Director of Regulatory Affairs at 635-778-4350 or email to: [email protected] from 8am to 5pm.

Distribution

Worldwide Distribution: US (nationwide) to states of: AL, AR, AZ, CA, CT, FL, HI, IA, IL, IN, KY, MA, MI, MN, MS, NC, NE, NJ, NM, NY, OH, OR, PA, RI, SC, SD, TN, TX, WA, WI, VA; and OUS (foreign) to countries of: Australia, Chile, China, France, Germany, India, Ireland, Italy, Japan, Korea, Mexico, Netherlands, Philippines, Poland, Puerto Rico, Spain, Switzerland, Taiwan, United Kingdom, Vietnam.

Quantity

14,789 (8651 US; 6138 OUS)