FDA Recall Open, Classified

Brand Name: Leica Biosystems Cryostats Product Name: CM1950 Model/Catalog Number: 1491950C4US Software Version: firmware V3.01

Recall: Z-1247-2025 · Initiated January 30, 2025

Recall

Recall Number
Z-1247-2025
Event Number
96220
Firm
LEICA BIOSYSTEMS NUSSLOCH GMBH Heidelberger Str. 17-19 Nussloch Germany
FEI Number
3002179321
Product Code
IDP
Status
Open, Classified
Root Cause
Under Investigation by firm
Initiated
January 30, 2025
Posted
February 27, 2025

Description

Brand Name: Leica Biosystems Cryostats Product Name: CM1950 Model/Catalog Number: 1491950C4US Software Version: firmware V3.01

Reason

Firm became aware of software issues related to most recent firmware update which could result in loss of patient tissue.

Action

On January 30, 2025, the firm notified customers through Email of the recall. Customers were provided with immediate actions to take to mitigate the risk of damaging the tissue sample. Among the firm's recommendations was to have a backup instrument if available. The firm's Service organization will contact customers in the coming weeks to organize the upgrade of affected instruments.

Distribution

Worldwide - US Nationwide distribution in the states of Idaho and Ohio. The countries of France, Australia, Belgium, China, Germany, Spain, UK, Hong Kong, Ireland, Italy, Japan, Netherlands, Norway, Poland, Singapore.

Quantity

2