Oertli 1. 23G Laser Probe OS4 Flex-Tip VK401113 2. 23G Laser Probe OS4 Illuminated Flex-Tip VK401213 3. 23G Laser Probe OS4 Intuitive Exendable VK401323 4. 25G Laser Probe OS4 Flex-Tip VK401115 5. 25G Laser Probe OS4 Illuminated Flex-Tip VK401215 6. 25G Laser Probe OS4 Intuitive Exendable VK401325 7. 27G Laser Probe OS4 Intuitive Exendable VK401327
Recall
- Recall Number
- Z-1246-2026
- Event Number
- 98255
- Firm
- Vortex Surgical Inc.
- FEI Number
- 3013002167
- Product Code
- HQB
- Status
- Open, Classified
- Root Cause
- Employee error
- Initiated
- December 16, 2025
- Posted
- February 4, 2026
- Address
- 4 Research Park Dr, Ste 124, Saint Charles, MO, 63304-5639
Description
Oertli 1. 23G Laser Probe OS4 Flex-Tip VK401113 2. 23G Laser Probe OS4 Illuminated Flex-Tip VK401213 3. 23G Laser Probe OS4 Intuitive Exendable VK401323 4. 25G Laser Probe OS4 Flex-Tip VK401115 5. 25G Laser Probe OS4 Illuminated Flex-Tip VK401215 6. 25G Laser Probe OS4 Intuitive Exendable VK401325 7. 27G Laser Probe OS4 Intuitive Exendable VK401327
There may be voids located in the seal of Tyvek pouches associated with identified catalog numbers and lots. A compromised sterile barrier has potential of bioburden contamination which could lead to infection.
On December 16, 2025, the firm began notifying customers via "URGENT: MEDICAL DEVICE RECALL" letters. Customers were instructed to discontinue use of affected product and to return product for Vortex Surgical to evaluate the seals; use UPS account 23623Y to ship unused products to Vortex Surgical Inc for replacement or refund at: Attn: Andrew Ritts Recall Tyvek Seal, Vortex Surgical Inc., 4 Research Park Dr., Suite 124, St. Charles, MO 63304; complete and return the Acknowledgement and Receipt Form via mail or email to: [email protected]. Actions have been implemented to improve training of individuals associated with sealing and inspecting seals to prevent reoccurrence. The sealer associated with the issue is being updated to include a guard to prevent reoccurrence. If you have any questions, contact the Director of Regulatory Affairs at 635-778-4350 or email to: [email protected] from 8am to 5pm.
Worldwide Distribution: US (nationwide) to states of: AL, AR, AZ, CA, CT, FL, HI, IA, IL, IN, KY, MA, MI, MN, MS, NC, NE, NJ, NM, NY, OH, OR, PA, RI, SC, SD, TN, TX, WA, WI, VA; and OUS (foreign) to countries of: Australia, Chile, China, France, Germany, India, Ireland, Italy, Japan, Korea, Mexico, Netherlands, Philippines, Poland, Puerto Rico, Spain, Switzerland, Taiwan, United Kingdom, Vietnam.
14,789 (8651 US; 6138 OUS)