FDA Recall
Terminated
SOMATOM Edge Plus, Model Number 10267000 Product Usage: Computed tomography systems intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data.
Recall: Z-1246-2019
·
Initiated February 14, 2019
Recall
- Recall Number
- Z-1246-2019
- Event Number
- 82227
- Firm
- Siemens Medical Solutions USA, Inc
- FEI Number
- 2240869
- Product Code
- JAK
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- February 14, 2019
- Terminated
- May 1, 2020
- Address
- 40 Liberty Blvd, Malvern, PA, 19355-1418
Description
SOMATOM Edge Plus, Model Number 10267000 Product Usage: Computed tomography systems intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data.
Reason
Potential risk of scans being aborted when using the optional Dual Spiral Dual Engery mode due to a software issue found in syngo CT VB10A for the Siemens Healthineers CT scanners with a single X-ray tube.
Action
Siemens issued letter dated 2/14/19 providing reason for recall and a user workaround in the Customer Safety Advisory Notice (CSAN) CT006/19/S. The recall issue will be permanently resolved by software update syngo CT VB10A_SP1. Contact the service organization 1-800-888-7436.
Distribution
US Nationwide Distribution
Quantity
53 total units distributed in U.S.