FDA Recall Terminated

ENSEAL Laparoscopic 5 mm Diameter Tissue Sealer G2 35 cm Length Curved Jaw, Model # NSLG2C35 and ENSEAL Laparoscopic 5 mm Diameter Tissue Sealer G2 45 cm Length Curved Jaw, Model # NSLG2C45. Ethicon Endo-Surgery, LLC Guaynabo, Puerto Rico 00969 The ENSEAL G2 Tissue Sealers are indicated for bipolar coagulation and mechanical transection of tissue during laparoscopic and open procedures. The devices are bipolar electrosurgical instruments for use with an electrosurgical generator. They are intended for use during open or laparoscopic, general and gynecological surgery to cut and seal vessels, and to cut, grasp and dissect tissue during surgery. Indications for use include open and laparoscopic, general and gynecological surgical procedures (including urologic, thoracic, plastic and reconstructive, bowel resections, hysterectomies, cholecystectomies, gall bladder procedures, Nissen fundoplication, adhesiolysis, oophorectomies, etc.), or any procedure where vessel ligation (cutting and sealing), tissue grasping and dissection is performed. The devices can be used on vessels up to (and including) 7 mm and bundles as large as will fit in the jaws of the instruments.

Recall: Z-1245-2012 · Initiated January 26, 2012

Recall

Recall Number
Z-1245-2012
Event Number
61314
Firm
Ethicon Endo-Surgery Inc
FEI Number
1527736
Product Code
LFL
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
January 26, 2012
Posted
March 21, 2012
Terminated
February 15, 2013
Address
4545 Creek Rd, Cincinnati, OH, 45242-2803

Description

ENSEAL Laparoscopic 5 mm Diameter Tissue Sealer G2 35 cm Length Curved Jaw, Model # NSLG2C35 and ENSEAL Laparoscopic 5 mm Diameter Tissue Sealer G2 45 cm Length Curved Jaw, Model # NSLG2C45. Ethicon Endo-Surgery, LLC Guaynabo, Puerto Rico 00969 The ENSEAL G2 Tissue Sealers are indicated for bipolar coagulation and mechanical transection of tissue during laparoscopic and open procedures. The devices are bipolar electrosurgical instruments for use with an electrosurgical generator. They are intended for use during open or laparoscopic, general and gynecological surgery to cut and seal vessels, and to cut, grasp and dissect tissue during surgery. Indications for use include open and laparoscopic, general and gynecological surgical procedures (including urologic, thoracic, plastic and reconstructive, bowel resections, hysterectomies, cholecystectomies, gall bladder procedures, Nissen fundoplication, adhesiolysis, oophorectomies, etc.), or any procedure where vessel ligation (cutting and sealing), tissue grasping and dissection is performed. The devices can be used on vessels up to (and including) 7 mm and bundles as large as will fit in the jaws of the instruments.

Reason

Ethicon Endo-Surgery initiated a voluntary global recall for specific production lots of ENSEAL G2 Curved and Straight Tissue Sealers due to two potential issues that may occur related to the activation button: (i) Continuous Activation: The ENSEAL device/system may continue to activate when the activation button is released prior to reaching the end of cycle (tone 3). (ii) No Activation

Action

Ethicon Endo-Surgery sent an Urgent Medical Device Recall letter dated February 10, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to immediately stop use of the affected product, fill out the attached Business Reply Form (BRF) and return to and return the affected product to: Ethicon Endo-Surgery ATT: ENSEAL G2 Recall 4545 Greek Road Cincinnati, Ohio 45242 Customers were instructed to choose one of the following response options: Return the Business Reply Form to their sales representative Call 1-800-873-3636, Option 6 Fax the BRF to 1-513-337-4138 For any questions regarding this recall call 513-337-3419.

Distribution

Worldwide Distribution - USA including AZ, CA, CT, FL, IL, IN, KS, KY, MA, MI, MO, NC, NY, OH, PA, TN & TX and the countries of Arab Emirates, Austria, Belgium, France, Germany, Great Britain, Italy, Jordan, Kingdom of Saudi Arabia, Lebanon, Portugal, Slovenia, Sweden and Turkey.

Quantity

633 pieces