FDA Recall Terminated

IMPAX 6.2.1 System, Picture Archiving and Communications system (PAC), used in the acceptance, transfer, display, storage and digital processing of medical images.

Recall: Z-1245-2008 · Initiated June 28, 2007

Recall

Recall Number
Z-1245-2008
Event Number
46386
Firm
AGFA Corp.
FEI Number
3001236302
Product Code
LLZ
Status
Terminated
Root Cause
Software design
Initiated
June 28, 2007
Posted
September 11, 2008
Terminated
March 3, 2011
Address
10 S Academy St, Greenville, SC, 29601-2632

Description

IMPAX 6.2.1 System, Picture Archiving and Communications system (PAC), used in the acceptance, transfer, display, storage and digital processing of medical images.

Reason

Text and Images may not be synchronized.

Action

AGFA notified all consignees via registered letter on/about 06/28/2007 informing them of the potential problem and requesting the accounts ensure the visual synchronization indicators in the Text and Image Areas are used to verify the synchronization status of the Text and Image Areas. A FAX Back form was requested with each letter to ensure someone at the account acknowledges the information was received and understood. If a FAX Back form is not received, the account will be contacted by telephone. A Mandatory Service Bulletin was deployed to correct the problem in the software.

Distribution

U.S. Nationwide and Canada

Quantity

154 units