FDA Recall Terminated

daVinci S Surgical system, patient side cart Auxiliary power board (APB) (printed circuit board assembly), endoscopic instrument control system.

Recall: Z-1245-2007 · Initiated May 14, 2007

Recall

Recall Number
Z-1245-2007
Event Number
38301
Firm
Intuitive Surgical, Inc.
FEI Number
3001675293
Product Code
NAY
Status
Terminated
Root Cause
Process control
Initiated
May 14, 2007
Posted
September 26, 2007
Terminated
December 3, 2007
Address
950 Kifer Rd, Sunnyvale, CA, 94086-5206

Description

daVinci S Surgical system, patient side cart Auxiliary power board (APB) (printed circuit board assembly), endoscopic instrument control system.

Reason

Product may malfunction and fail to start up on AC power.

Action

As of 9/12/2007, the firm has conducted a field inspection of any involved units via service visit. No other notice is planned or indicated by the firm's submission. The firm feels that the risk associated with this issue is minimal, and therefore, any advanced notification is unneccessary. If any system failed one or more of the inspection points, its APB (3510404-04) was replaced with a verified APB (351040-04 or 351040-05).

Distribution

Worldwide, including USA, France, China, Japan, Mexico, Spain, Italy, Belgium and Singapore.

Quantity

38 units