WEB1000 Versions 3.1, 4.1 and 5.1, Picture Archiving and Communications systems (PACs), used in the acceptance, transfer, display, storage and digital processing of medical images.

Recall: Z-1244-2008 · Initiated May 14, 2007

Recall

Recall Number
Z-1244-2008
Event Number
46384
Firm
AGFA Corp.
FEI Number
3001236302
Product Code
LLZ
Status
Terminated
Root Cause
Software change control
Initiated
May 14, 2007
Posted
September 11, 2008
Terminated
December 9, 2010
Address
10 S Academy St, Greenville, SC, 29601-2632

Description

WEB1000 Versions 3.1, 4.1 and 5.1, Picture Archiving and Communications systems (PACs), used in the acceptance, transfer, display, storage and digital processing of medical images.

Reason

Java version 1.6x (JAVA 6) is not compatible with AGFA IMPAX Administration/Service Tools. Due to a Software Mismatch data can be lost.

Action

Agfa Healthcare contacted all consignees on 5/14/2007 and by registered mail on 06/07/2007. The registered Urgent Field Safety Notice letters requested the accounts take the following steps immediately: 1. Turn off automatic JAVA updates to avoid this update being applied in the background. 2. Do not manually update JAVA on their station-even for other applications. 3. If JAVA 1.6 has already been applied, do not use Service/Admin Tools. They were instructed to contact their IT department or Agfa support for assistance in backing out of this version to apply the required version for the Agfa software. A Mandatory Service Bulletin will inform and assist affected account in removing JAVA 1.6 if it has been applied and turning off the automatic updates via the JAVA Plug-in Control Panel.

Distribution

U.S. Nationwide and Canada.

Quantity

946 units (total)