IMPAX 4.1, IMPAX 4.5, IMPAX 5.0, IMPAX 5.1 IMPAX 5.2, IMPAX 6.0, IMPAX 6.2; Picture Archiving and Communications systems (PACs), used in the acceptance, transfer, display, storage and digital processing of medical images.
Recall
- Recall Number
- Z-1243-2008
- Event Number
- 46384
- Firm
- AGFA Corp.
- FEI Number
- 3001236302
- Product Code
- LLZ
- Status
- Terminated
- Root Cause
- Software change control
- Initiated
- May 14, 2007
- Posted
- September 11, 2008
- Terminated
- December 9, 2010
- Address
- 10 S Academy St, Greenville, SC, 29601-2632
Description
IMPAX 4.1, IMPAX 4.5, IMPAX 5.0, IMPAX 5.1 IMPAX 5.2, IMPAX 6.0, IMPAX 6.2; Picture Archiving and Communications systems (PACs), used in the acceptance, transfer, display, storage and digital processing of medical images.
Java version 1.6x (JAVA 6) is not compatible with AGFA IMPAX Administration/Service Tools. Due to a Software Mismatch data can be lost.
Agfa Healthcare contacted all consignees on 5/14/2007 and by registered mail on 06/07/2007. The registered Urgent Field Safety Notice letters requested the accounts take the following steps immediately: 1. Turn off automatic JAVA updates to avoid this update being applied in the background. 2. Do not manually update JAVA on their station-even for other applications. 3. If JAVA 1.6 has already been applied, do not use Service/Admin Tools. They were instructed to contact their IT department or Agfa support for assistance in backing out of this version to apply the required version for the Agfa software. A Mandatory Service Bulletin will inform and assist affected account in removing JAVA 1.6 if it has been applied and turning off the automatic updates via the JAVA Plug-in Control Panel.
U.S. Nationwide and Canada.
946 units (total for both types of software)