FDA Recall Open, Classified

Optima MR360, NMRI system

Recall: Z-1241-2025 · Initiated January 31, 2025

Recall

Recall Number
Z-1241-2025
Event Number
96311
Firm
GE Healthcare (China) Co., Ltd. 1F-5F BLD 1, No.1 Tongji South Road, Beijing Economic-Technological Development Area Beijing China
FEI Number
3008284470
Product Code
LNH
Status
Open, Classified
Root Cause
Software design
Initiated
January 31, 2025
Posted
February 28, 2025

Description

Optima MR360, NMRI system

Reason

For certain MR systems, a scan can be resumed following a Power Monitor trip when the "6-minute average SAR" is above the limit indicated on the "SAR Display". The user interface may incorrectly enable a scan to proceed earlier than intended. If resumed by the operator, such a scan can result in potential excessive tissue heating, especially if the patient is in contact with a conductive material or has an implant.

Action

On January 31, 2025, the firm notified customers through Urgent Medical Device Correction letters. Customers were provided with instructions for continued use of their MR systems. All potential users should be made aware of this safety notification and recommended actions. If you have any questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative.

Distribution

Domestic US distribution nationwide. International distribution worldwide.

Quantity

33