FDA Recall Terminated

Agfa HeartLab Cardiovascular DICOMstore Version 2.04.40.00 software, Heartlab, Inc., Westerly, Rhode Island 02891. The software is used in the Agfa HeartLab Cardiovascular System.

Recall: Z-1241-2008 · Initiated June 21, 2007

Recall

Recall Number
Z-1241-2008
Event Number
46381
Firm
AGFA Corp.
FEI Number
3001236302
Product Code
LLZ
Status
Terminated
Root Cause
Software design
Initiated
June 21, 2007
Posted
June 27, 2008
Terminated
August 4, 2008
Address
10 S Academy St, Greenville, SC, 29601-2632

Description

Agfa HeartLab Cardiovascular DICOMstore Version 2.04.40.00 software, Heartlab, Inc., Westerly, Rhode Island 02891. The software is used in the Agfa HeartLab Cardiovascular System.

Reason

MIsidentification: The electrocardiogram (ECG) data of one patient is misidentified as the data of another patient.

Action

On 6/21-26/2007, AGFA Healthcare notified, via telephone, the affected consignees of the situation. On 7/2/2007, affected consignees were sent, via registered mail, an "Urgent Medical Device Correction" letter instructing them to: a) remove the affected software DICOMstore 2.04.40.00 and replace it with the prior software version, DICOMstore 2.04.34.02; and b) review patient ECG reports to determine if any patients were actually affected and to notify the firm. AGFA Healthcare also provided a customer service contact number (877) 777-2432.

Distribution

Worldwide Distributution - USA, including states of AZ, CA, DC, FL, IL, IN, KY, MA, ME, MN, MS, OH, OK, OR, NH, PA, SC, TN, TX, WI, and country of Canada.

Quantity

43 units