FDA Recall Terminated

Kimberly-Clark/Ballard medical R2 triangle shaped non-radiolucent multifunction electrode products for pediatrics.

Recall: Z-1241-06 · Initiated May 18, 2006

Recall

Recall Number
Z-1241-06
Event Number
35559
Firm
Kimberly-Clark Corporation
FEI Number
1000015879
Product Code
MKJ
Status
Terminated
Root Cause
Device Design
Initiated
May 18, 2006
Posted
July 28, 2006
Terminated
May 14, 2007
Address
1400 Holcomb Bridge Rd, Roswell, GA, 30076

Description

Kimberly-Clark/Ballard medical R2 triangle shaped non-radiolucent multifunction electrode products for pediatrics.

Reason

The tinfoil in the triangle electrodes has the potential to develop cracks which may hinder electrical flow and product performance during defibrillation.

Action

Consignees were first contacted by phone with follow up letter sent via fax or mail on May 18, 2006. International consignees will be notified through the firm's regional offices.

Distribution

Worldwide ( including USA, Switzerland, Germany, UK, Canada, Australia, Netherlands, Italy, Greece, Spain, Hong Kong, Hungary, Denmark, Singapore, New Zealand, Ireland, Korea)

Quantity

95,639 electrode sets