Siemens ADVIA Centaur Enhance Estradiol Test Code eE2 Catalog Number: 10490889, 10491445, 10697757 Product Usage: For in vitro diagnostic use in the quantitative determination of estradiol in human serum and plasma (heparinized and EDTA) using the ADVIA Centaur, ADVIA Centaur XP, ADVIA Centaur XPT and ADVIA Centaur CP systems.
Recall
- Recall Number
- Z-1240-2016
- Event Number
- 73052
- Firm
- Siemens Healthcare Diagnostics, Inc
- FEI Number
- 1219913
- Product Code
- CHP
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- January 13, 2016
- Terminated
- January 8, 2018
- Address
- 333 Coney St, East Walpole, MA, 02032-1516
Description
Siemens ADVIA Centaur Enhance Estradiol Test Code eE2 Catalog Number: 10490889, 10491445, 10697757 Product Usage: For in vitro diagnostic use in the quantitative determination of estradiol in human serum and plasma (heparinized and EDTA) using the ADVIA Centaur, ADVIA Centaur XP, ADVIA Centaur XPT and ADVIA Centaur CP systems.
The drug fulvestrant (Faslodex) may cause falsely elevated estradiol results.
Siemens issued an Urgent Medical Device Correction (UMDC) to affected Siemens Healthcare Diagnostics customers in the United States on January 13, 2016 to be delivered to customers on January 14, 2016 via Federal Express. An Urgent Field Safety Notice (UFSN) was sent out for distribution to all customers outside the United States on January 13, 2016. The letters identified affected product, problem and the actions to be taken. Customers may continue to use the estradiol assays to report patient results however, the assays should not be used for patients being treated with fulvestrant. Customers where instructed to complete and return the Correction Effectiveness Check attached. For questions contact your Siemens Customer Care Center or your local Siemens technical support representative.
Worldwide Distribution US Nationwide , AR, AU, AU, BD, BO, BR, CA, CH, CL, CN, CO, DO, EC, EG, HK, ID, IE, IL, IN, JP, KR, LK, LY, MX, MY, NI, NP, NZ, PE, PH, PY, SA, SG, SV, TH, TW, UY, VE, VN, and ZA.
33604 units