Access Ostase Reagent, Part Number: 37300, subsequent product code: JIS. The Access Ostase Assay is used for the quantitative measurement of bone alkaline phosphatase (BAP), an indicator of osteoblastic activity, in human serum and plasma. This device is intended to be used as an aid in the management of postmenopausal osteoporosis and Pagets disease.
Recall
- Recall Number
- Z-1240-2012
- Event Number
- 61214
- Firm
- Beckman Coulter Inc.
- FEI Number
- 2050012
- Product Code
- CIN
- Status
- Terminated
- Root Cause
- Process design
- Initiated
- November 10, 2011
- Posted
- March 21, 2012
- Terminated
- June 26, 2012
- Address
- 250 S Kraemer Blvd, Brea, CA, 92821-6232
Description
Access Ostase Reagent, Part Number: 37300, subsequent product code: JIS. The Access Ostase Assay is used for the quantitative measurement of bone alkaline phosphatase (BAP), an indicator of osteoblastic activity, in human serum and plasma. This device is intended to be used as an aid in the management of postmenopausal osteoporosis and Pagets disease.
The recall was initiated because Beckman confirmed that the Access Ostase Calibrator and QC lots do not meet expiration date claims. The Access Ostase Reagent Kit lots identified were quality control tested and released using Access Ostase Calibrator and QC lots that contain the implicated raw material.
Beckman Coulter sent an "URGENT PRODUCT CORRECTION" letter dated November 9, 2011 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. The letter instructs customers to discontinue use and discard all remaining inventory. In addition, a Response Form was enclosed for customers to complete and return. Contact Beckman's Customer Support Center at (800) 854-3633 for questions regarding this notice.
Worldwide Distribution-USA (nationwide) and the countries of Algeria, Argentina, Australia, Austria, Belgium, Canada, China, Colombia, Czech Republic, Denmark, France, Georgia, Germany, Greece, Hong Kong, Hungary, India,Ireland, Italy, Japan, Jordan, Lebanon, Liechtenstein, Mexico, Morocco, Netherlands, New Zealand, Panama, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Turkey, United Kingdom, and Venezuela.
8228 units total (1193 units in US)