FDA Recall Terminated

BCI Autocorr Plus Pulse Oximeter/ECG/Respiration Monitor, Catalog Numbers, 3404001 (including S) and 3404002 (including E,F, G, and S) Smiths Medical- a part of Smiths Group plc

Recall: Z-1240-2007 · Initiated July 18, 2007

Recall

Recall Number
Z-1240-2007
Event Number
38472
Firm
Smiths Medical PM, Inc.
FEI Number
2183502
Product Code
DQA
Status
Terminated
Root Cause
Other
Initiated
July 18, 2007
Posted
September 21, 2007
Terminated
October 31, 2008
Address
N7w22025 Johnson Dr, Waukesha, WI, 53186-1856

Description

BCI Autocorr Plus Pulse Oximeter/ECG/Respiration Monitor, Catalog Numbers, 3404001 (including S) and 3404002 (including E,F, G, and S) Smiths Medical- a part of Smiths Group plc

Reason

Phantom pacer spikes caused by a damaged resistor.

Action

Customers will be notified via a Customer Information Bulletin CIB Tracking Number: 07-CIB 12, July 18,2007 requesting them to evalute the functionality of their monitor (s). If the monitor does not exhibit any false pace detect makers when pace detect is enabled and respiration is disabled, no further action is required. If it does exhibit false pace detect they will be instructed to contact the Smiths Medical PM, Inc service department. At that time Technical Worksheet, 07-TW02, will be provided. The consignee will need to verify they are capable of perfoming the Technical Worksheet provided. The Smiths Medical PM. Inc. Service Manger will veirfy if the consignee is capable of performing the rework.

Distribution

World wide: USA including states of KY, MS, PA, LA, and PR and countries of Poland, Mexico, Kuwait, Taiwan, Singapore, Hon Kong, India, Bolivia, Colombia, Chile, Argentia, Brunei Darssalam, Israel, Greece, Romania, Germany, The Netherlands, France, Great Britain, Spain, Brazil, Russia, China.

Quantity

178