SIGNA Creator, SIGNA Explorer, NMRI systems
Recall
- Recall Number
- Z-1239-2025
- Event Number
- 96311
- Firm
- GE Healthcare (China) Co., Ltd. 1F-5F BLD 1, No.1 Tongji South Road, Beijing Economic-Technological Development Area Beijing China
- FEI Number
- 3008284470
- Product Code
- LNH
- Status
- Open, Classified
- Root Cause
- Software design
- Initiated
- January 31, 2025
- Posted
- February 28, 2025
Description
SIGNA Creator, SIGNA Explorer, NMRI systems
For certain MR systems, a scan can be resumed following a Power Monitor trip when the "6-minute average SAR" is above the limit indicated on the "SAR Display". The user interface may incorrectly enable a scan to proceed earlier than intended. If resumed by the operator, such a scan can result in potential excessive tissue heating, especially if the patient is in contact with a conductive material or has an implant.
On January 31, 2025, the firm notified customers through Urgent Medical Device Correction letters. Customers were provided with instructions for continued use of their MR systems. All potential users should be made aware of this safety notification and recommended actions. If you have any questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative.
Domestic US distribution nationwide. International distribution worldwide.
2538