International Technidyne Corp. (ITC) AVOXimeter 1000E Oximeter, Catalog Numbers: AVOX1000E-110 and AVOX1000E-220
Recall
- Recall Number
- Z-1238-2008
- Event Number
- 46773
- Firm
- International Technidyne Corp.
- FEI Number
- 1000526865
- Product Code
- DQA
- Status
- Terminated
- Root Cause
- Process design
- Initiated
- February 13, 2008
- Posted
- August 6, 2008
- Terminated
- July 24, 2009
- Address
- 6-8 Olsen Ave, Edison, NJ, 08820-2419
Description
International Technidyne Corp. (ITC) AVOXimeter 1000E Oximeter, Catalog Numbers: AVOX1000E-110 and AVOX1000E-220
Incorrect reading: ITC discovered that instruments built or repaired since approximately April 2007 were inadvertently left with the diagnostic mode enabled when shipped to customers.
ITC sent Urgent - Medical Device Field Recall - ITC No. 08-001 letters dated 2/08/2008 to hospitals and/or medical facilities identified as having affected instruments. The letter instructed users not to use the instrument if the diagnostic mode is enabled. ITC will reset affected devices and return them to the customers.
Worldwide Distribution: USA, United Arab Emirates, Malaysia, Saudi Arabia, India, Germany, United Kingdom, Greece, Romania, Singapore, ROC, Pakistan, Netherlands, Korea, France, Belgium, Switzerland, Canada, Thailand, Slovenia, Turkey and Australia.
164 US; 39 International