Axiom Artis Systems operating with A100 Generator
Recall
- Recall Number
- Z-1237-2010
- Event Number
- 54999
- Firm
- Siemens Medical Solutions USA, Inc
- FEI Number
- 2240869
- Product Code
- MQB
- Status
- Terminated
- Root Cause
- Nonconforming Material/Component
- Initiated
- March 4, 2010
- Posted
- April 5, 2010
- Terminated
- July 5, 2012
- Address
- 51 Valley Stream Pkwy, Malvern, PA, 19355
Description
Axiom Artis Systems operating with A100 Generator
Due to a contact problem at the low voltage power supply of the generator controller, in rare cases where a user has invoked radiation, it may occur that no radiation is released. Under certain circumstances, radiation is possible by repeated release of the x-ray footswitch or button. To remedy the issue described above, a revised cabling has been installed in all potentially affected syste
The recalling firm issued a Customer Safety Advisory to affected customers via Update Instructions AX006/10/S. The letter informs customers of the potential issues and provides additional information on the implemented resolutions. For additional information, please contact the firm at 888-826-9702.
The product was shipped to medical facilities nationwide.
1373 units