FDA Recall Terminated

Axiom Artis Systems operating with A100 Generator

Recall: Z-1237-2010 · Initiated March 4, 2010

Recall

Recall Number
Z-1237-2010
Event Number
54999
Firm
Siemens Medical Solutions USA, Inc
FEI Number
2240869
Product Code
MQB
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
March 4, 2010
Posted
April 5, 2010
Terminated
July 5, 2012
Address
51 Valley Stream Pkwy, Malvern, PA, 19355

Description

Axiom Artis Systems operating with A100 Generator

Reason

Due to a contact problem at the low voltage power supply of the generator controller, in rare cases where a user has invoked radiation, it may occur that no radiation is released. Under certain circumstances, radiation is possible by repeated release of the x-ray footswitch or button. To remedy the issue described above, a revised cabling has been installed in all potentially affected syste

Action

The recalling firm issued a Customer Safety Advisory to affected customers via Update Instructions AX006/10/S. The letter informs customers of the potential issues and provides additional information on the implemented resolutions. For additional information, please contact the firm at 888-826-9702.

Distribution

The product was shipped to medical facilities nationwide.

Quantity

1373 units