FDA Recall Terminated

Siemens Atellica IM aTG (100 test kit) Siemens Material Number (SMN): 10995461

Recall: Z-1235-2019 · Initiated December 7, 2018

Recall

Recall Number
Z-1235-2019
Event Number
82521
Firm
Siemens Healthcare Diagnostics, Inc
FEI Number
1219913
Product Code
JZO
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
December 7, 2018
Terminated
April 27, 2020
Address
333 Coney St, East Walpole, MA, 02032-1516

Description

Siemens Atellica IM aTG (100 test kit) Siemens Material Number (SMN): 10995461

Reason

Positive bias anti-thyroglobulin with ADVIA Centaur aTG kit lots ending in 316 and lower and Atellica IM aTG kit lots ending in 317 and lower when compared to the standardization to World Health Organization (WHO) Reference Preparation MRC 65/93 stated in the Instructions for Use (IFU).

Action

Siemens issued Urgent Medical Device Corrections, to Siemens Healthcare Diagnostics customers in the United States on December 7, 2018 delivered to customers on December 10, 2018. Urgent Field Safety Notices sent out to the country organizations on December 7, 2018 for distribution to customers outside the United States. These letters inform customers that Siemens Healthcare Diagnostics identified a bias with the affected lots of ADVIA Centaur aTG and Atellica IM aTG and action to take and resolution of the issue with new lots of the reagents. Questions contact your Siemens Customer Care Center or your local Siemens technical support representative.

Distribution

Nationwide Foreign: Belgium¿¿¿¿¿¿¿¿ Croatia¿¿¿¿¿¿¿¿ Czech¿Republic¿ Denmark¿¿¿¿¿¿¿¿ France¿¿¿¿¿¿¿¿¿ Germany¿¿¿¿¿¿¿¿ Greece¿¿¿¿¿¿¿¿¿ Ireland¿¿¿¿¿¿¿¿ Italy¿¿¿¿¿¿¿¿¿¿ Poland¿¿¿¿¿¿¿¿¿ Portugal¿¿¿¿¿¿¿ Slovakia¿¿¿¿¿¿¿ Spain¿¿¿¿¿¿¿¿¿¿ Switzerland¿¿¿¿ Turkey¿¿¿¿¿¿¿ BOGOTA QUARRY BAY, HONG KONG YANGON SANTIAGO DE CHILE MONTEVIDEO BUENOS AIRES LIMA JOINVILLE, SC, BRAZIL Selangor Darul Ehsan Dhaka Ha Noi City Gurgaon, Haryana Hanoi-Hoangmai ASUNCION SHANGHAI Seoul 120 SINGAPORE RIYADH 11411 TAIPEI R.O.C. QUITO HUNTINGWOOD, NEW SOUTH WALES ZAPOPAN Ho Chi Minh-District 3 Huaykwang, Bangkok RIYADH SAN JOSE CHIBA LAVAL HO CHI MINH CITY MANAGUA Gurgaon, Haryana

Quantity

1489 kits