FDA Recall Terminated

Zip 8M Wound Closure Device; for wounds up to 8 cm in length; Product number PS9080; Rx only, Sterile Indicated for use during and after skin incision procedures to approximate skin and hold together skin edges until healing can take place.

Recall: Z-1235-2015 · Initiated February 5, 2015

Recall

Recall Number
Z-1235-2015
Event Number
70466
Firm
Zipline Medical
FEI Number
3009673389
Product Code
KGX
Status
Terminated
Root Cause
Packaging process control
Initiated
February 5, 2015
Posted
March 2, 2015
Terminated
April 10, 2015
Address
747 Camden Ave, Ste A, Campbell, CA, 95008-4147

Description

Zip 8M Wound Closure Device; for wounds up to 8 cm in length; Product number PS9080; Rx only, Sterile Indicated for use during and after skin incision procedures to approximate skin and hold together skin edges until healing can take place.

Reason

The manufacturers seal may not be present on the Nylon pouch resulting in lack of assurance for sterility.

Action

Zipline sent an Urgent Voluntary product Recall letters to customers on February 5, 2015 by e-mail with possible follow up on site by firm representatives. The letter identified the affected product, problem and actions to be taken. Customers are asked to quarantine inventory immediately and examine the affected lots for complete pouch seals. Any product found with an incomplete seal should be destroyed or returned to ZipLine Medical. Please complete the enclosed response form as soon as possible, even if you no longer have stock of the lot numbers subject to this correction. This information is required to reconcile the inventory. Please contact Customer Service at 1-888-326-8999 to coordinate any product returns, replacement stock, or credit notes.

Distribution

Worldwide Distribution - US Nationwide and the countries of Panama, Tehran, Ireland, Sweden, Japan, Hong Kong, Finland, Pakistan, Taiwan, Malaysia.

Quantity

3109 devices, total all model numbers