FDA Recall Open, Classified

Socrates 38 Aspiration Catheter 127 cm length REF SC038-127-001 The Socrates Aspiration System with a compatible suction pump is intended for use in the revascularization of patients with acute ischemic stroke

Recall: Z-1234-2025 · Initiated January 2, 2025

Recall

Recall Number
Z-1234-2025
Event Number
96158
Firm
Scientia Vascular, Inc.
FEI Number
3009546466
Product Code
NRY
Status
Open, Classified
Root Cause
Under Investigation by firm
Initiated
January 2, 2025
Posted
February 26, 2025
Address
2460 S 3270 W, West Valley City, UT, 84119-1116

Description

Socrates 38 Aspiration Catheter 127 cm length REF SC038-127-001 The Socrates Aspiration System with a compatible suction pump is intended for use in the revascularization of patients with acute ischemic stroke

Reason

Due to manufacturing non-conformance that resulted in channels in the seal on the packaging pouch. As a result, this could potential create a breach in the sterile barrier.

Action

On 01/02/2025, the firm initially notified customer via email an "URGENT: MEDICAL DEVICE RECALL" Letter informing them that through internal evaluation of finished goods of the Socrates 38 Aspiration Catheter, one lot of pouches containing the 127 cm Socrates 38 Aspiration Catheter was identified with pouches that had a gap in the sterile barrier seal. This was a result of the pouch not being centered with the sealing bar during the sealing process. On 01/10/2025, The firm sent an updated letter informing customers of additional product lots were added to the Recall. Customers are instructed to: Segregate Recalled Product. Immediately remove all impacted (see list above) Socrates 38 Catheter(s) from their inventory (regardless of its location) and segregate this product in a secure location for return to Scientia Vascular, Inc. In situations where a device has already been used in a patient procedure, closely monitor patients for signs and symptoms of post-surgical infection. Customer Service will be available to provide additional information and provide replacement orders at [email protected] or (888)385-9016; Monday through Friday, 8:00 AM to 4:30 PM, Mountain Time.

Distribution

U..S. Nationwide distribution in the states of AL, AZ, FL, GA, HI, IL, IN, KS, MN, NC, NJ, NY, PA, SD, TX, VA, WI and WV.

Quantity

120 devices