FDA Recall Terminated

ACE Trochanteric Nail System ATN LAG SCREW 60MM STERILE ATN LAG SCREW 65MM STERILE ATN LAG SCREW 70MM STERILE ATN LAG SCREW 75MM STERILE ATN LAG SCREW 80MM STERILE ATN LAG SCREW 85MM STERILE ATN LAG SCREW 90MM STERILE ATN LAG SCREW 95MM STERILE ATN LAG SCREW 100MM STERILE ATN LAG SCREW 105MM STERILE ATN LAG SCREW 110MM STERILE ATN LAG SCREW 115MM STERILE ATN LAG SCREW 120MM STERILE ATN LAG SCREW 125MM STERILE ATN LAG SCREW 130MM STERILE The Trochanteric Nail System is intended to treat stable and unstable proximal fractures of the femur including pertrochanteric fractures, intertrochanteric fractures, high subtrochanteric fractures and combinations of these fractures. The Trochanteric Long Nail System is additionally indicated to treat pertrochanteric fractures associated with shaft fractures, pathologic fractures in osteoporotic bone of the trochanteric and diaphyseal areas, long subtrochanteric fractures, ipsilateral femoral fractures, proximal or distal nonunions and malunions and revision procedures.

Recall: Z-1234-2018 · Initiated December 20, 2017

Recall

Recall Number
Z-1234-2018
Event Number
79408
Firm
Zimmer Biomet, Inc.
FEI Number
1825034
Product Code
HSB
Status
Terminated
Root Cause
Device Design
Initiated
December 20, 2017
Posted
March 14, 2018
Terminated
April 22, 2020
Address
56 E Bell Dr, Warsaw, IN, 46582-6989

Description

ACE Trochanteric Nail System ATN LAG SCREW 60MM STERILE ATN LAG SCREW 65MM STERILE ATN LAG SCREW 70MM STERILE ATN LAG SCREW 75MM STERILE ATN LAG SCREW 80MM STERILE ATN LAG SCREW 85MM STERILE ATN LAG SCREW 90MM STERILE ATN LAG SCREW 95MM STERILE ATN LAG SCREW 100MM STERILE ATN LAG SCREW 105MM STERILE ATN LAG SCREW 110MM STERILE ATN LAG SCREW 115MM STERILE ATN LAG SCREW 120MM STERILE ATN LAG SCREW 125MM STERILE ATN LAG SCREW 130MM STERILE The Trochanteric Nail System is intended to treat stable and unstable proximal fractures of the femur including pertrochanteric fractures, intertrochanteric fractures, high subtrochanteric fractures and combinations of these fractures. The Trochanteric Long Nail System is additionally indicated to treat pertrochanteric fractures associated with shaft fractures, pathologic fractures in osteoporotic bone of the trochanteric and diaphyseal areas, long subtrochanteric fractures, ipsilateral femoral fractures, proximal or distal nonunions and malunions and revision procedures.

Reason

Zimmer Biomet is conducting a medical device recall for the ACE Trochanteric Nail (ATN) system after a Device Performance Review (DPR) failed clinical analysis for survivorship pursuant to the acceptance criteria for the DPR protocol for the ATN system.

Action

On December 22, 2017 an URGENT MEDICAL DEVICE RECALL letter was issued to customers with a complete list of products subject to recall and requesting that the customer review the notification and notify all affected personnel of the recall. If there is affected product at the facility it is requested that these units be quarantined and returned to a sales representative for return to the firm. Questions or concerns can be directed to 574-371-3071.

Distribution

USA (nationwide)

Quantity

3454