FDA Recall
Terminated
Dimension Vista Enzymatic Creatinine (ECREA) reagent, SMN 10700444
Recall: Z-1233-2020
·
Initiated December 18, 2019
Recall
- Recall Number
- Z-1233-2020
- Event Number
- 84612
- Firm
- Siemens Healthcare Diagnostics, Inc.
- FEI Number
- 2432235
- Product Code
- JFY
- Status
- Terminated
- Root Cause
- Under Investigation by firm
- Initiated
- December 18, 2019
- Posted
- January 24, 2020
- Terminated
- June 25, 2020
- Address
- 511 Benedict Ave, Tarrytown, NY, 10591-5005
Description
Dimension Vista Enzymatic Creatinine (ECREA) reagent, SMN 10700444
Reason
There is a potential for falsely depressed creatinine results for patients on phenindione therapy when using the enzymatic methodology. Interference has not been observed with the creatinine Jaffe methodology.
Action
Siemens Healthcare Diagnostics initiated a customer communication for this issue via letter beginning December 18, 2019.
Distribution
US Nationwide
Quantity
11,725 units