FDA Recall Terminated

cobas p 512 pre-analytical system Pre-analytical sample handling that includes de-capping, aliquoting and sorting of samples for analysis.

Recall: Z-1233-2016 · Initiated February 23, 2016

Recall

Recall Number
Z-1233-2016
Event Number
73292
Firm
Roche Diagnostics Operations, Inc.
FEI Number
1823260
Product Code
JQP
Status
Terminated
Root Cause
Software design
Initiated
February 23, 2016
Terminated
January 11, 2018
Address
9115 Hague Rd, Indianapolis, IN, 46256-1025

Description

cobas p 512 pre-analytical system Pre-analytical sample handling that includes de-capping, aliquoting and sorting of samples for analysis.

Reason

Due to a false triggering or detection of the lifting gripper READY signal, sample tubes are not correctly placed back in the Rack Tube Transport (RTT) after the decapping process. Therefore, open sample tubes can be dropped in the cobas p 512, spilling the sample material.

Action

Roche Diagnostics sent an "Urgent Medical Device Correction" letter via UPS Ground (receipt required) on February 23, 2016, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. To prevent this malfunction, a software patch has been developed and validated. The installation of this software patch is mandatory and will be performed by a Roche Field Engineering Specialist (FES). Your Roche FES will contact you to schedule the installation of the software patch. Until the software patch is implemented, Roche recommends that instrument operators monitor their systems for the occurrence of Error 140 Gripper has lost tube. If this error occurs, all components in contact with a spill must have extra cleaning. Refer to the cobas p 512 pre-analytical instrument Operators Manual Version 1.5. Please contact the Roche Support Network Customer Support Center, 24 hours a day, seven days a week at 1-800-428-2336 for further questions.

Distribution

US Distribution including Puerto Rico and to the states of :TX, OH, TN, AZ, WA, MI and GA

Quantity

35 Units