FDA Recall Terminated

Titan" Stabilizer Attachment and Titan TM 360 Stabilizer Attachment

Recall: Z-1232-2017 · Initiated January 5, 2015

Recall

Recall Number
Z-1232-2017
Event Number
76346
Firm
Terumo Cardiovascular Systems Corporation
FEI Number
3000204839
Product Code
MWS
Status
Terminated
Root Cause
Other
Initiated
January 5, 2015
Posted
February 22, 2017
Terminated
February 23, 2017
Address
125 Blue Ball Rd, Elkton, MD, 21921-5315

Description

Titan" Stabilizer Attachment and Titan TM 360 Stabilizer Attachment

Reason

Terumo Cardiovascular Systems (Terumo CVS) is voluntarily recalling certain lots of Titan" Stabilizers and Titan" 360 Stabilizers because the Suction Tubing Clip may break when it is being attached to a Terumo CVS Hercules Stabilizing Arm. Although the clip is for convenience purposes and is not required for device operation, there is a potential for broken Suction Tubing Clip fragments to fall into the surgical site.

Action

A recall notification is being forwarded to all users and consignees who have received product. The notification of this voluntary recall will be sent to the following consignee in each facility:+ Director of Operating Room Services Proposed Level of Effectiveness Checks: Terumo CVS will verify that 100% of all letters are delivered to affected consignees

Distribution

The product was distributed domestically and also to foreign consignees. But not to government agency's or Canada.

Quantity

1463