FDA Recall Open, Classified

PRISMAFLEX HF1000 SET, Product Code 107142; Dialyzer, High Permeability With Or Without Sealed Dialysate System

Recall: Z-1230-2026 · Initiated January 6, 2026

Recall

Recall Number
Z-1230-2026
Event Number
98180
Firm
VANTIVE US HEALTHCARE LLC
FEI Number
3038195011
Product Code
KDI
Status
Open, Classified
Root Cause
Component design/selection
Initiated
January 6, 2026
Posted
February 2, 2026
Address
510 Lake Cook Rd, Deerfield, IL, 60015-4964

Description

PRISMAFLEX HF1000 SET, Product Code 107142; Dialyzer, High Permeability With Or Without Sealed Dialysate System

Reason

There is a potential dislodgement of the Prismaflex Set deaeration chamber from the Prismaflex Control Unit.

Action

Vantive issued an Urgent Medical Device Correction notice to its consignees on 01/06/2026 via USPS first class mail. The notice explained the issue, hazard involved, and requested the following: 1. You may continue to perform therapy using the Prismaflex sets listed above. Please monitor the deaeration chamber during therapy to ensure it remains in an upright position within the holder. 2. Vantive is aware of customers who have experienced this issue and attempted to further secure the deaeration chamber in an upright position. If attempting to secure the deaeration chamber in an upright position, please consider the following: . Ensure there are no kinks in the tubing and that the deaeration chamber remains visible. . If the Prismaflex Control Unit issues an Air in Blood alarm, check if air is present. If there is no air present, check if the deaeration chamber is dislodged from the holder and ensure there are no clots present. 1. If no dislodgement, please follow the operator s manual instructions for Air in Blood alarm. 2. In case of dislodgement with clotting, follow the associated instructions within the Prismaflex Control Unit operator s manual to stop the therapy and replace the set. 3. Only in case of dislodgement without clotting or any other alarms, if the chamber is secured in an upright position, follow the associated instructions within the Prismaflex Control Unit operator s manual to proceed with treatment. Facilities who further distributed the products were directed to notify those departments or facilities. Dealers, wholesalers, distributor/resellers, or original equipment manufacturers (OEM) who further distributed the products were directed to notify their customers. For general questions regarding this communication, please contact Vantive Customer Care at 833-542-CRRT (2778). Vantive Customer Care is available 24 hours a day, seven days a week.

Distribution

US Nationwide distribution.

Quantity

115,112 units