FDA Recall Open, Classified

VITROS 4600 Chemistry System Catalog Nos. 6802445 (regular) and 6900440 (refurbished)

Recall: Z-1230-2024 · Initiated January 26, 2024

Recall

Recall Number
Z-1230-2024
Event Number
93927
Firm
Ortho-Clinical Diagnostics, Inc.
FEI Number
1000136573
Product Code
JJE
Status
Open, Classified
Root Cause
Software Design Change
Initiated
January 26, 2024
Posted
March 1, 2024
Address
100 Indigo Creek Dr, Rochester, NY, 14626-5101

Description

VITROS 4600 Chemistry System Catalog Nos. 6802445 (regular) and 6900440 (refurbished)

Reason

A software defect affecting VITROS Systems running VITROS Software Versions 3.8.0 and 3.8.1 causes the Configurable Westgard Rules to not report as expected. When a new control is defined, configurable Westgard rules, if selected, are not flagged or reported with the applicable QC results if the statistical criteria is met. This may result in acceptance of QC results that do not meet acceptance criteria, which may cause erroneous patient results to be reported.

Action

On January 26, 2024, the firm issued a customer letter to all customers who have V3.8.0 or V3.8.1 installed on their VITROS Analyzer. Customers were informed of the issue and instructed to follow the instructions when using the VITROS System onboard QC package to use the configurable Westgard Rules. The software will be updated to mitigate this issue within the VITROS Software QC feature. The estimated availability is 2Q2024. If you have further questions, please contact Global Services Organization at 1-800-421-3311.

Distribution

Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Norway, Portugal, Russia, Singapore, Spain, Sweden, The Netherlands, United Kingdom.

Quantity

15 US; 43 OUS