FDA Recall Terminated

ECMO Reinforced Dual Lumen Cannula, Sterile EO, Rx Only, REF: VV13F

Recall: Z-1230-2019 · Initiated January 17, 2017

Recall

Recall Number
Z-1230-2019
Event Number
82512
Firm
OriGen Biomedical, Inc.
FEI Number
1000160256
Product Code
DWF
Status
Terminated
Root Cause
Process control
Initiated
January 17, 2017
Posted
June 4, 2019
Terminated
July 27, 2021
Address
7000 Burleson Rd, Bldg D, Austin, TX, 78744-3202

Description

ECMO Reinforced Dual Lumen Cannula, Sterile EO, Rx Only, REF: VV13F

Reason

Clamping the clear tube very close to the blue hub may cause a localized adhesive failure in the bond between the clear tube and the molded hub. Forcefully rotating the clamp while it is in place may potentially cause an immediate disconnect of the clear tube or lead to a later fatigue failure of the adhesive bond, with subsequent clear tube disconnect from the hub.

Action

On 01/17/2017, Technical Bulletin 17.01 was mailed to customers. Customers were informed to clamp all sizes of catheters only in the middle. The following contact information was provided: US: 1-512-474-7278, and Germany: +49-7475-915591.

Distribution

US: NY, AB, AR, FL, MO, NC, GA, LA, CO, AL, MA, CA, MI, WI, TX, DC, OH, CT, TN, IN, OR, IL, MN, QC, SC, KY, OK, AZ, UT, RI, NJ, SD, PA, MS, NM. Foreign (OUS): UAE, Saudi Arabia, Colombia, UK, Netherlands, Italy, Czech Republic, Germany, India, Japan, Canada, Thailand, Sweden, Belgium, Spain, Poland, South Africa, Chile

Quantity

145