ECMO Reinforced Dual Lumen Cannula, Sterile EO, Rx Only, REF: VV32F
Recall
- Recall Number
- Z-1229-2019
- Event Number
- 82512
- Firm
- OriGen Biomedical, Inc.
- FEI Number
- 1000160256
- Product Code
- DWF
- Status
- Terminated
- Root Cause
- Process control
- Initiated
- January 17, 2017
- Posted
- June 4, 2019
- Terminated
- July 27, 2021
- Address
- 7000 Burleson Rd, Bldg D, Austin, TX, 78744-3202
Description
ECMO Reinforced Dual Lumen Cannula, Sterile EO, Rx Only, REF: VV32F
Clamping the clear tube very close to the blue hub may cause a localized adhesive failure in the bond between the clear tube and the molded hub. Forcefully rotating the clamp while it is in place may potentially cause an immediate disconnect of the clear tube or lead to a later fatigue failure of the adhesive bond, with subsequent clear tube disconnect from the hub.
On 01/17/2017, Technical Bulletin 17.01 was mailed to customers. Customers were informed to clamp all sizes of catheters only in the middle. The following contact information was provided: US: 1-512-474-7278, and Germany: +49-7475-915591.
US: NY, AB, AR, FL, MO, NC, GA, LA, CO, AL, MA, CA, MI, WI, TX, DC, OH, CT, TN, IN, OR, IL, MN, QC, SC, KY, OK, AZ, UT, RI, NJ, SD, PA, MS, NM. Foreign (OUS): UAE, Saudi Arabia, Colombia, UK, Netherlands, Italy, Czech Republic, Germany, India, Japan, Canada, Thailand, Sweden, Belgium, Spain, Poland, South Africa, Chile
66