ACE Trochanteric Nail System TROCH NAIL ST 11X180 125DG STRL TROCH NAIL ST 11X180 130DG STRL TROCH NAIL ST 11X180 135DG STRL The Trochanteric Nail System is intended to treat stable and unstable proximal fractures of the femur including pertrochanteric fractures, intertrochanteric fractures, high subtrochanteric fractures and combinations of these fractures. The Trochanteric Long Nail System is additionally indicated to treat pertrochanteric fractures associated with shaft fractures, pathologic fractures in osteoporotic bone of the trochanteric and diaphyseal areas, long subtrochanteric fractures, ipsilateral femoral fractures, proximal or distal nonunions and malunions and revision procedures.
Recall
- Recall Number
- Z-1229-2018
- Event Number
- 79408
- Firm
- Zimmer Biomet, Inc.
- FEI Number
- 1825034
- Product Code
- HSB
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- December 20, 2017
- Posted
- March 14, 2018
- Terminated
- April 22, 2020
- Address
- 56 E Bell Dr, Warsaw, IN, 46582-6989
Description
ACE Trochanteric Nail System TROCH NAIL ST 11X180 125DG STRL TROCH NAIL ST 11X180 130DG STRL TROCH NAIL ST 11X180 135DG STRL The Trochanteric Nail System is intended to treat stable and unstable proximal fractures of the femur including pertrochanteric fractures, intertrochanteric fractures, high subtrochanteric fractures and combinations of these fractures. The Trochanteric Long Nail System is additionally indicated to treat pertrochanteric fractures associated with shaft fractures, pathologic fractures in osteoporotic bone of the trochanteric and diaphyseal areas, long subtrochanteric fractures, ipsilateral femoral fractures, proximal or distal nonunions and malunions and revision procedures.
Zimmer Biomet is conducting a medical device recall for the ACE Trochanteric Nail (ATN) system after a Device Performance Review (DPR) failed clinical analysis for survivorship pursuant to the acceptance criteria for the DPR protocol for the ATN system.
On December 22, 2017 an URGENT MEDICAL DEVICE RECALL letter was issued to customers with a complete list of products subject to recall and requesting that the customer review the notification and notify all affected personnel of the recall. If there is affected product at the facility it is requested that these units be quarantined and returned to a sales representative for return to the firm. Questions or concerns can be directed to 574-371-3071.
USA (nationwide)
3454