FDA Recall Terminated

HeartMate II System Controller Product Usage: HeartMate II is a mechanical device that circulates blood throughout the body when the heart is too weak to pump blood adequately on its own. It is sometimes called a a heart pump or VAD. HeartMate II is a small implantable LVAD. HeartMate II attaches to the heart and is designed to assist or take over the pumping function of the patients left ventricle, the main pumping chamber of the heart.

Recall: Z-1228-2014 · Initiated March 4, 2014

Recall

Recall Number
Z-1228-2014
Event Number
67639
Firm
Thoratec Corporation
FEI Number
2916596
Product Code
DSQ
Status
Terminated
Root Cause
Device Design
Initiated
March 4, 2014
Posted
March 21, 2014
Terminated
August 27, 2015
Address
6035 Stoneridge Dr, Pleasanton, CA, 94588-3270

Description

HeartMate II System Controller Product Usage: HeartMate II is a mechanical device that circulates blood throughout the body when the heart is too weak to pump blood adequately on its own. It is sometimes called a a heart pump or VAD. HeartMate II is a small implantable LVAD. HeartMate II attaches to the heart and is designed to assist or take over the pumping function of the patients left ventricle, the main pumping chamber of the heart.

Reason

Discovery of serious injuries and deaths associated with the process of changing from a primary System controller to their back-up System controller in patients using the Pocket System controller model.

Action

Thoratec sent an Urgent Medical Device Correction letter dated March 4, 2014 to all affected customers. The letter was sent via e-mail. All distributors outside of the USA and Canada to whom affected product has been distributed have been notified of this field correction notice and have been provided copies of the Customer Letter to forward to their customers. The letter identified the affected product , problem and actions to be taken. Customers were requested to complete and sign the attached Acknowledgement Form and return it to Thoratec via fax (925-847-8571) or e-mail a scanned copy to [email protected]. For questions call 1-800-528-2577.

Distribution

Worldwide distribution: US (nationwide) including states of: AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NJ, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, and WI; and countries of: Austria, Belgium, Czech Republic, France, Germany, Ireland, Israel, Italy, Lebanon, Lithuania, Poland, Saudi Arabia, Singapore, Slovenia, Sweden, Switzerland, The Netherlands, Turkey, and United Kingdom.

Quantity

4120