FDA Recall Open, Classified

Back up batteries in Ventilators labeled as the following: a. CARESCAPE R860 Ventilators (K142679) b. ENGSTROM (K111116) c. ENGSTROM Carestation (K111116) d. ENGSTROM PRO (K111116)

Recall: Z-1226-2022 · Initiated April 18, 2022

Recall

Recall Number
Z-1226-2022
Event Number
90076
Firm
GE Healthcare, LLC
FEI Number
2126677
Product Code
CBK
Status
Open, Classified
Root Cause
Under Investigation by firm
Initiated
April 18, 2022
Posted
June 6, 2022
Address
3000 N Grandview Blvd, Waukesha, WI, 53188-1615

Description

Back up batteries in Ventilators labeled as the following: a. CARESCAPE R860 Ventilators (K142679) b. ENGSTROM (K111116) c. ENGSTROM Carestation (K111116) d. ENGSTROM PRO (K111116)

Reason

Insufficient battery backup power resulting in premature shutdown of the ventilator when not connected to AC mains power supply.

Action

The firm, GE Healthcare, sent an "URGENT MEDICAL DEVICE CORRECTION" letter the consignees via mail on 04/18/2022. The letter describes the product, problem and actions to be taken. The customers instructed to take the following actions: 1. You can continue to use the affected ventilators while the ventilator is connected to an AC mains power source that is supported by backup emergency power. 2. If it is absolutely necessary to use the ventilator by relying on the battery (such as during required transport where there are no other options), ensure you follow standard clinical practice of having a readily accessible means of appropriate alternative ventilation (for example a bag-valve system) and personnel with the capability to administer this alternative means at all times. 3. Immediately after receiving this communication, perform the Battery Performance Test as described in Appendix A. Replace the batteries when necessary, before patient use. 4. When not in patient use, it is recommended that the device always remains connected to the AC mains power source to prevent battery discharge and degradation, even when in storage. 5. It is recommended that the Battery Performance Test be completed every three months as described in Appendix A. 6. If the device has been in storage for over three months, perform the Battery Performance Test as described in Appendix A prior to use. 7. The backup batteries must be replaced at a minimum every three years. 8. Complete the attached Medical Device Notification Acknowledgement Response form and send to [email protected] If you have any questions or concerns regarding this notification, please contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.

Distribution

Worldwide Distribution: US (nationwide) to AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NH, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI and OUS (internationally) to Australia, Austria, Bangladesh, Belgium, Brazil, Bulgaria, Canada, Chile, China, Columbia, Cote D'Ivoire, Mexico, Uruguay, United Kingdom, United Arab Emirates, Turkey, Thailand, Tanzania, Sweden, Spain, Singapore, Saudi Arabia, Russia, Romania, Portugal, Poland, Philippines, Panama, Omar, Norway, Nigeria, Nicaragua, New Zealand, Netherlands, Malaysia, Latvia, Kuwait, Korea, Kenya, Kazakhstan, Japan, Italy, Ireland, Indonesia, India, Hungary, Germany, France, Finland, Estonia, Egypt, Ecuador, Denmark, Czech Republic.

Quantity

34,940 ventilators