GE Inspection Technologies x|argos 160D system. Cabinet x-ray system.

Recall: Z-1225-2015 · Initiated January 13, 2015

Recall

Recall Number
Z-1225-2015
Event Number
70478
FEI Number
2519904
Product Code
RCE
Status
Terminated
Root Cause
Device Design
Initiated
January 13, 2015
Posted
March 20, 2015
Terminated
August 17, 2015
Address
Rd, GE Inspection Technologies, LP 50 Industrial Park, Lewistown, PA, 17044-9312

Description

GE Inspection Technologies x|argos 160D system. Cabinet x-ray system.

Reason

It was discovered that the upper left corner-shaped lead shield in the window in one of the doors installed had lost adhesion and fall out of position, which could lead to gaps in shielding.

Action

GE Inspection Technologies, Planned Action: 1. GE will notify customers of the affected cabinet x-ray systems through a Technical Information Letter. GE notified its affected customers via email on January 13, 2015, via letter dated January 14, 2015, and via telephone on January 19, 2015. 2.GE will install an interim insert within 30 days of receipt of this letter. The insert will cover any existing or interim gaps in the original lead shielding on the door. 3. GE will replace the doors of affected cabinet x-ray systems no later than 180 day of receipt of this letter. 4. These corrections will be made free of charge. The Technical Information Letter used for notification of all dealers and purchasers is to be made within 15 working days of receipt of this letter For further questions, please call (717) 447-1278

Distribution

US Distribution to the states of OH and PA.

Quantity

2