BMSI "MicroJack" PN 672-510000 patient interface electrode connection device used during electroencephalography (EEG) /Epilepsy Monitoring evaluations with various EEG monitoring systems. The MicroJack is used with BMSI Models 2000, 3000, 4000, 5000 & 6000, and is used with the BMSI kit upgraded with NicoletOne. Nicolet Biomedical, Inc.

Recall: Z-1225-2007 · Initiated July 17, 2007

Recall

Recall Number
Z-1225-2007
Event Number
38424
Firm
Nicolet Biomedical Div of Viasys Healthcare
FEI Number
3010611950
Product Code
GWQ
Status
Terminated
Root Cause
Other
Initiated
July 17, 2007
Posted
September 20, 2007
Terminated
December 10, 2011
Address
5225 Verona Rd, Madison, WI, 53711-4497

Description

BMSI "MicroJack" PN 672-510000 patient interface electrode connection device used during electroencephalography (EEG) /Epilepsy Monitoring evaluations with various EEG monitoring systems. The MicroJack is used with BMSI Models 2000, 3000, 4000, 5000 & 6000, and is used with the BMSI kit upgraded with NicoletOne. Nicolet Biomedical, Inc.

Reason

The BMSI "MicroJack" patient interface electrode connection device used with the EEG monitor systems was found to be mis-wired. A complainant alleged that during depth electrode EEG study of a patient in preparation for brain surgery there was an observation that was perceived as non-logical data output.

Action

VIASYS will initiate a field correction for units that have been distributed. The correction consists of testing the units and replacing those that have been found to contain the defect. Method of communication will be by certiified mail to identify consignees and receipt of the communication will be tracked. Letters "Viasys HealthCare Urgent Medical Device Field Correction (Recall)" were sent 7/17/07 to both domestic and international customers.

Distribution

Worldwide, including USA, ISRAEL, LITHUANIA, RUSSIA, FRANCE, IRELAND, UNITED KINGDOM , SPAIN, YEMEN, JAPAN, TAIWAN, KOREA, CHINA, SAUDI ARABIA, HONG KONG, INDIA, COLOMBIA, BRAZIL, GERMANY, CANADA, DENMARK, ITALY, MEXICO, and THAILAND.

Quantity

2102