Hamamatsu Corporation, Bridgewater NJ 08807 Microfocus X-ray Source L8601 series, L9121 series (90kV type MICROFOCUS X-RAY SOURCE), L7901 series, and L8031 series (100kV type MICROFOCUS x-ray source
Recall
- Recall Number
- Z-1224-2012
- Event Number
- 60991
- Firm
- Hamamatsu Corporation
- FEI Number
- 1000121818
- Product Code
- RBS
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- March 30, 2011
- Posted
- March 29, 2012
- Terminated
- January 29, 2018
- Address
- 360 Foothill Rd, Bridgewater, NJ, 08807
Description
Hamamatsu Corporation, Bridgewater NJ 08807 Microfocus X-ray Source L8601 series, L9121 series (90kV type MICROFOCUS X-RAY SOURCE), L7901 series, and L8031 series (100kV type MICROFOCUS x-ray source
It was discovered that the Micro Focus X-ray Source could be experiencing software related problems that would allow the unit to exceed the maximum power setting that was set by the operator.
Hamamtsu Corporation sent a Warning letter dated December 22, 2011, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. CDRH approves your CAP under the following conditions: 1. Hamamatsu is able to work with your manufacturer customers to assure the revised manual and new label are provided to your customers' end users. 2. Hamamatsu informs CDRH if you receive information which indicates this control defect has occurred. Relevant information includes direct communication from customers or through observation of excessive wear on returned x-ray tubes. For questions regarding this recall call 908-231-0960, ext 2300.
Worldwide Distribution-- USA (nationwide) and the country of Canada.