Vitros Chemistry *** GLU (glucose) Slides *** 5 cartridges each containing 60 slides for a total of 300 tests per box *** REF 170 7801 *** Ortho-Clinical Diagnostics Inc., 100 Indigo Creek Drive, Rochester, NY 14626. [NOTE: product has an 18 month shelf life.]

Recall: Z-1224-03 · Initiated July 28, 2003

Recall

Recall Number
Z-1224-03
Event Number
27035
Firm
Ortho-Clinical Diagnostics
FEI Number
1000136573
Product Code
CGA
Status
Terminated
Root Cause
Other
Initiated
July 28, 2003
Posted
September 11, 2003
Terminated
November 24, 2003
Address
100 Indigo Creek Drive, Rochester, NY, 14650-0881

Description

Vitros Chemistry *** GLU (glucose) Slides *** 5 cartridges each containing 60 slides for a total of 300 tests per box *** REF 170 7801 *** Ortho-Clinical Diagnostics Inc., 100 Indigo Creek Drive, Rochester, NY 14626. [NOTE: product has an 18 month shelf life.]

Reason

Results of glucose measurements on sodium fluoride/potassium oxalate specimens can be negatively biased when using Vitros GLU and GLU DT slides stored in the refrigerator. GLU slides are processed by high volume chemistry systems typically used in hospital laboratories.

Action

GLU Slides: Letters dated between 7/28/03 and 8/11/03 provide revised storage instructions. Customers using NaF/KOx blood collection tubes and have VITROS GLU slides that have been refrigerated for more than four months will receive replacement product. Those customers are instructed to continue to use the slides only until replacement product arrives. GLU DT Slides: Letters dated between 7/28/03 and 8/11/03 with the same instructions as for GLU slides.

Distribution

Nationwide and worldwide distribution to distributors and end-users.

Quantity

487,912 boxes