Ikaria, INOmax DSIR (Delivery System), Model 10007. Nitric oxide delivery system for use with ventilators.
Recall
- Recall Number
- Z-1223-2015
- Event Number
- 70395
- Firm
- INO Therapeutics (dba Ikaria)
- FEI Number
- 3004531588
- Product Code
- MRN
- Status
- Terminated
- Root Cause
- Process change control
- Initiated
- January 14, 2015
- Posted
- February 27, 2015
- Terminated
- April 15, 2015
- Address
- 2902 Dairy Dr, Madison, WI, 53718-3809
Description
Ikaria, INOmax DSIR (Delivery System), Model 10007. Nitric oxide delivery system for use with ventilators.
Potential delivery failure alarm condition. INOmax DSIR with software version 2.0.4 and a certain variant of the monitor display may trigger a delivery failure alarm when the display brightness is set to its lowest level. If this condition occurs, therapy will be interrupted.
Consignees were sent on 1/14/2015 an Ikaria "Urgent Field Correction Action" letter dated January 14, 2015. The letter described the problem and the product involved in the recall. The letter provided a number of recommendations to the consginees in regard to use and return of the product. For questions or adverse events they can contact Ikaria Customer Care at 1-877-KNOW-INO (1-877-566-9466) 24 hours a day, 7 days a week.
Distribution US nationwide, including AZ, CA, CT, FL, ID, LA, MD, MI, MS, NH, NJ, NV, NY, OK, PA, TX, UT, VA, WV.
45 units