FDA Recall Terminated

Leksell GammaPlan, Model 5.34. Intended to be used for planning the dosimetry of treatments in stereotactic radiation therapy.

Recall: Z-1223-2011 · Initiated October 1, 2005

Recall

Recall Number
Z-1223-2011
Event Number
57607
Firm
Elekta, Inc.
FEI Number
1037831
Product Code
IWB
Status
Terminated
Root Cause
Other
Initiated
October 1, 2005
Posted
February 14, 2011
Terminated
March 2, 2011
Address
4775 Peachtree Industrial Blvd, Bldg 300 # 300, Norcross, GA, 30092-3011

Description

Leksell GammaPlan, Model 5.34. Intended to be used for planning the dosimetry of treatments in stereotactic radiation therapy.

Reason

Investigation found that if a user accidentally selects the wrong image/tube position images can be displayed flipped.

Action

Product bulletin Leksell GammaPlan - "Leksell GammaPlan version 5.34" (dated October 2005) was sent to affected Customers as notification of the problem. Field Change Order FCO-DCO-05-0006 (dated November 22, 2005) and Technical Notes Leksell GammaPlan Leksell SurgiPlan Important Notice - "Potential issue using flipped angiographic images" (dated October 2005) was sent to Elekta Service Engineers as notification of the problem. The letters described the issue and the FCO asked the Service Engineers to confirm their receipt of the documentation. The Technical Notes also discussed image visualization requirements and listed recommendations.

Distribution

Worldwide Distribution

Quantity

272 units