COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conjunction with a cyanide-free lytic agent for counting and sizing blood cells on all UniCel DxH Coulter Cellular Analysis Systems. For use as a rinsing agent on the UniCel DxH Slidemaker Stainers. For use as a diluent in conjunction with DxH ECO Cell Lyse for counting and sizing blood cells. For use as a rinsing agent.
Recall
- Recall Number
- Z-1221-2026
- Event Number
- 97992
- Firm
- Beckman Coulter, Inc.
- FEI Number
- 3000203293
- Product Code
- GIF
- Status
- Open, Classified
- Root Cause
- Under Investigation by firm
- Initiated
- November 10, 2025
- Posted
- January 30, 2026
- Address
- 11800 Sw 147th Ave, Miami, FL, 33196-2500
Description
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conjunction with a cyanide-free lytic agent for counting and sizing blood cells on all UniCel DxH Coulter Cellular Analysis Systems. For use as a rinsing agent on the UniCel DxH Slidemaker Stainers. For use as a diluent in conjunction with DxH ECO Cell Lyse for counting and sizing blood cells. For use as a rinsing agent.
Due to specific lots of diluents contributing to elevated platelet Daily Checks background counts cycle with use with analyzers.
During the week of 11/10/2025, the firm sent via postal and e-mail an "URGENT MEDICAL DEVICE RECALL" Letter to customers informing them that Beckman Coulter has confirmed that specific diluent lot numbers of COULTER DxH Diluent and DxH ECO Diluent may contribute to elevated platelet (PLT) Daily Checks background counts cycle when used with the UniCel DxH 600/690T/800/900 analyzers. Customers are instructed to: "Check their COULTER DxH Diluent and DxH ECO Diluent inventory against the provided affected lot numbers. "If their lot numbers are NOT affected, continue routine operation and no further action is required. "If their lot numbers are affected, perform a Daily Checks cycle. o Each container must be individually verified through Daily Checks before use according to the instructions below. o If Daily Checks pass, continue routine operation and no further action is required, on the active diluent container(s), currently in use. "If the Daily Checks cycle does not pass, follow the instruction provide in the Recall Letter. For questions or further assistance - email [email protected]
Worldwide - U.S. Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DC, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PA, Puerto Rico, RI, SC, SD, TN, TX, UT, VA, WA, WI, and WV. The countries of Argentina, Canada, Chile, Mexico, Saint Kitts and Nevis, Trinidad and Tobago, and Viet Nam.
76,098 units (US-61843 and OUS-14255)