FDA Recall Terminated

Shutter HX SP5, a component of the Leica Point Scanning Confocal Microscope Models TCS SP5 and TCS SP5 X. Laboratory instrument.

Recall: Z-1221-2010 · Initiated November 12, 2009

Recall

Recall Number
Z-1221-2010
Event Number
54165
Firm
Leica Microsystems, Inc.
FEI Number
3000210133
Product Code
REM
Status
Terminated
Root Cause
Radiation Control for Health and Safety Act
Initiated
November 12, 2009
Posted
May 6, 2010
Terminated
May 11, 2010
Address
2345 Waukegan Rd, Bannockburn, IL, 60015-1515

Description

Shutter HX SP5, a component of the Leica Point Scanning Confocal Microscope Models TCS SP5 and TCS SP5 X. Laboratory instrument.

Reason

A manufacturing defect was identified that could potentially result in the laser shutter not closing as intended which could result in the shutter not completely closing disrupting the laser beam safely.

Action

A Service Note was sent to service organizations on November 12, 2009. The communication asked customers to check if they must exchange their shutters and send their planned exchange date by 2009-11-20 to [email protected]. Customers were also asked to check the serial numbers of the one-wire chips of the shutters. Then lastly, the were asked to exchange the shutters and confirm the exchange.

Distribution

Worldwide Distribution -- USA, Sweden, Japan, Hong Kong, Italy, the Netherlands, Singapore, UK, and Spain.

Quantity

15 microscopes and 8 spare part shutters