FDA Recall Terminated

ARROW 40cc Drive Line Kits, REF IAK-02692, Intra-aortic balloon catheter. The Intra-aortic balloon (IAB) is utilized for IAB counter-pulsation therapy, whereby balloon inflation in the aorta during diastole and deflation during systole increase blood supply to the heart muscle and decrease work of the left ventricle.

Recall: Z-1221-2009 · Initiated February 2, 2009

Recall

Recall Number
Z-1221-2009
Event Number
51080
Firm
Arrow International, Inc.
FEI Number
3010532612
Product Code
DSP
Status
Terminated
Root Cause
Process design
Initiated
February 2, 2009
Posted
April 16, 2009
Terminated
July 6, 2010
Address
9 Plymouth St, Everett, MA, 02149-1814

Description

ARROW 40cc Drive Line Kits, REF IAK-02692, Intra-aortic balloon catheter. The Intra-aortic balloon (IAB) is utilized for IAB counter-pulsation therapy, whereby balloon inflation in the aorta during diastole and deflation during systole increase blood supply to the heart muscle and decrease work of the left ventricle.

Reason

A fault in the connector of the pump tubing assembly may result in the volume setting on the pump defaulting to 2.5 ccs or 5 ccs (depending on IABP model) rather than the appropriate 30, 40, or 50 cc volume.

Action

An "Urgent Medical Device Recall" letter dated February 2, 2009 was issued to consignees via US Certified Mail to notify users of the problem. Users were instructed to check their inventory for the affected products, quarantine, and return the them to Arrow International, Inc. Customers were also instructed to follow the "Mitigation Directions" in circumstances where alternate Intra-Aortic Balloon (IAB) catheters/adaptors or other forms of therapy are not available (until receipt of available product). Canadian accounts were notified on February 9, 2009 by courier and International accounts notified on 6 February 2009 and 10 February 2009 via Registered Post/UPS. Direct questions to your local Sales Representative or the Arrow International Customer Service at 1-800-523-8446 (US) or 1-919-361-4062 (outside US). Press Release issued April 13, 2009 is posted to firm's webiste: http://www.arrowintl.com/iab_recall.asp.

Distribution

Worldwide Distribution: USA, Argentina, Australia, Taiwan, Singapore, New Zealand, India, Brazil, Belgium, Korea, Peru, Finland, and Denmark.

Quantity

90 units