FDA Recall Terminated

TriActiv System Procedure Kit. Balloon Protected Flush Extraction System 340 cm (for embolism protection). The kit includes AutoStream Flow Control, Flush Kit, ShieldWire (temporary occlusion balloon guidewire), and Balloon Inflation Syringe). The system is used with a standard 7F guide catheter using femoral approach.

Recall: Z-1220-06 · Initiated May 24, 2006

Recall

Recall Number
Z-1220-06
Event Number
35680
Firm
Kensey Nash Corp
FEI Number
1000307073
Product Code
NFA
Status
Terminated
Root Cause
Other
Initiated
May 24, 2006
Posted
July 12, 2006
Terminated
July 11, 2006
Address
735 Pennsylvania Dr, Exton, PA, 19341-1130

Description

TriActiv System Procedure Kit. Balloon Protected Flush Extraction System 340 cm (for embolism protection). The kit includes AutoStream Flow Control, Flush Kit, ShieldWire (temporary occlusion balloon guidewire), and Balloon Inflation Syringe). The system is used with a standard 7F guide catheter using femoral approach.

Reason

volume control knob not able to go up to larger diameters

Action

A letter was issued to the sales representative that have customers which received affected product, 5/24/2006. Sales representatives visited their sites within 3 days to determine if the affected product was still in inventory. If the product was still in inventory the sales representative provided the trained TriActiv System user(s) a Technical Bulletin that instructed the users how to correct the knob position.

Distribution

Nationwide. The products are shipped to hospitals in AL, DE, Fl, IL, IN, MI, PA, SC, TN and TX.

Quantity

10 kits