FDA Recall Terminated

PCS125 Swivel Collimator, Model F100; Manual Radiographic Collimator; Summit Industries, Inc., Chicago, Illinois

Recall: Z-1218-2007 · Initiated August 10, 2007

Recall

Recall Number
Z-1218-2007
Event Number
38512
Firm
Summit Industries Inc.
FEI Number
1450503
Product Code
IZX
Status
Terminated
Root Cause
Other
Initiated
August 10, 2007
Posted
August 29, 2007
Terminated
March 18, 2010
Address
2901 W Lawrence Ave, Chicago, IL, 60625-3621

Description

PCS125 Swivel Collimator, Model F100; Manual Radiographic Collimator; Summit Industries, Inc., Chicago, Illinois

Reason

The F100 Swivel Collimator may come loose from the tubestand and fall onto the x-ray table.

Action

The affected dealers were notified of the field correction recall by letter dated 8/3/07, informing them of the problem with the stripped screws and requesting that they notify their affected end user locations. The end user notification letters inform them of the problem and state that the dealer will be contacting them soon to arrange a time to visit the end user location and conduct the rework on the collimator, which should take less than 30 minutes. In the mean time the end user was instructed to examine their swivel collimator for looseness. If it is found to be loose, the end user was instructed to take the unit out of service and call the dealer.

Distribution

Worldwide, including USA, Australia, Cambodia, Canada, Chile, Colombia, Dominican Republic, Ecuador, France, Germany, Haiti, Mexico, Nicaragua, Singapore, and South Africa.

Quantity

3034 for all units