FDA Recall Terminated

KARL STORZ Photodynamic Diagnosis (PDD) D-Light C System, Instruction Manuals: PDD Telescopes, Tricam SL II, and PDD Camera Heads. Used as an optional accessory to white light cystoscopy when used in combination with the diagnostic imaging drug Cysview.

Recall: Z-1216-2013 · Initiated August 16, 2012

Recall

Recall Number
Z-1216-2013
Event Number
60555
Firm
Karl Storz Endoscopy America Inc
FEI Number
2020550
Product Code
OAY
Status
Terminated
Root Cause
Process control
Initiated
August 16, 2012
Posted
May 8, 2013
Terminated
May 8, 2013
Address
2151 East Grand Avenue, El Segundo, CA, 90245

Description

KARL STORZ Photodynamic Diagnosis (PDD) D-Light C System, Instruction Manuals: PDD Telescopes, Tricam SL II, and PDD Camera Heads. Used as an optional accessory to white light cystoscopy when used in combination with the diagnostic imaging drug Cysview.

Reason

The recall was initiated because Karl Storz has confirmed that the Photodynamic Diagnosis (PDD) D-Light C System instruction manuals have important safety-related changes.

Action

A Safety Alert letter, dated 8/16/12, was sent to all their customers to inform them that there is a labeling error in which important safety-related changes were made to the KARL STORZ Photodynamic Diagnosis (PDD) D-Light C System instruction manuals. The correct Ethylene Oxide (EtO) Sterilization set point for Relative Humidity should read, =70% RH. The correction applies to the PDD Camera Head instruction manual, PDD Rigid Telescopes instruction manual, and Tricam SL II instruction manual, the three component manuals indicated for use within the KARL STORZ Photodynamic Diagnosis (PDD) D-Light C System. The letter instructed the customers to place each of the three enclosed stickers, in the appropriate instruction manual. Customers are instructed to cover the erroneous information with the provided stickers as indicated. Customers with any questions regarding the proper use of the device, are instructed to contact Karl Storz's Technical Support Department at 800-421-0837, Ext. 5350, during office hours; or Ext. 7222, 24 hours/7 days a week for emergencies.

Distribution

Nationwide distribution: USA including states of: IL, CA, WV, TX, SC, WI, NY, PA, NE, NJ, MD, OH, IA, and NC.

Quantity

15 units