Philips EasyDiagnost Eleva DRF, Fluoroscopic x-ray system Software Release 3.0 and 3.1 Model: 706034 Product Usage: Radiological image processing system (computed radiography x-ray system)
Recall
- Recall Number
- Z-1216-2012
- Event Number
- 61204
- Firm
- Philips Healthcare Inc.
- FEI Number
- 1218950
- Product Code
- KPR
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- February 14, 2012
- Posted
- March 15, 2012
- Terminated
- September 23, 2020
- Address
- 3000 Minuteman Road, Andover, MA, 01810
Description
Philips EasyDiagnost Eleva DRF, Fluoroscopic x-ray system Software Release 3.0 and 3.1 Model: 706034 Product Usage: Radiological image processing system (computed radiography x-ray system)
Misinterpretation of Mirror Icon-Mirror icon on the bottom of each image could be misinterpreted as a RIGHT lead marker for side indication if the image is mirrored within a PACS system
Philips Healthcare sent an Urgent Medical Device Correction letter dated February 8, 2012 to all affected customers. The letter identified the affected product, problem description, hazard involved, actions to be taken by customer/user and actions planned by Philips. The letter contains important information for continue safe and proper use of the equipment. Philips will contact customers to schedule the installation of the updated software. For further information or support contact your local Philips representative at the Technical Support Line 866-767-2822.
Worldwide Distribution - USA Nationwide
60 units