FDA Recall Terminated

Philips EasyDiagnost Eleva DRF, Fluoroscopic x-ray system Software Release 3.0 and 3.1 Model: 706034 Product Usage: Radiological image processing system (computed radiography x-ray system)

Recall: Z-1216-2012 · Initiated February 14, 2012

Recall

Recall Number
Z-1216-2012
Event Number
61204
Firm
Philips Healthcare Inc.
FEI Number
1218950
Product Code
KPR
Status
Terminated
Root Cause
Software design
Initiated
February 14, 2012
Posted
March 15, 2012
Terminated
September 23, 2020
Address
3000 Minuteman Road, Andover, MA, 01810

Description

Philips EasyDiagnost Eleva DRF, Fluoroscopic x-ray system Software Release 3.0 and 3.1 Model: 706034 Product Usage: Radiological image processing system (computed radiography x-ray system)

Reason

Misinterpretation of Mirror Icon-Mirror icon on the bottom of each image could be misinterpreted as a RIGHT lead marker for side indication if the image is mirrored within a PACS system

Action

Philips Healthcare sent an Urgent Medical Device Correction letter dated February 8, 2012 to all affected customers. The letter identified the affected product, problem description, hazard involved, actions to be taken by customer/user and actions planned by Philips. The letter contains important information for continue safe and proper use of the equipment. Philips will contact customers to schedule the installation of the updated software. For further information or support contact your local Philips representative at the Technical Support Line 866-767-2822.

Distribution

Worldwide Distribution - USA Nationwide

Quantity

60 units