FDA Recall Terminated

Boston Scientific IQ Guide Wire. Catalog No. 38950-01J0, 185 cm, J-Tip, Single Pack. Sterile EO. Manufactured for: Boston Scientific Corporation, One Boston Scientific Place, Natick, MN 01760 USA. Boston Scientific Guide Wires are steerable guide wires available in a variety of lengths and diameters. The distal tip is shapeable, or, alternatively, is available in a preshaped ''J'' tip.

Recall: Z-1214-06 · Initiated May 30, 2006

Recall

Recall Number
Z-1214-06
Event Number
35549
Firm
Boston Scientific
FEI Number
3002095335
Product Code
DQX
Status
Terminated
Root Cause
Other
Initiated
May 30, 2006
Posted
July 12, 2006
Terminated
June 28, 2007
Address
1 Scimed Pl, Maple Grove, MN, 55311-1565

Description

Boston Scientific IQ Guide Wire. Catalog No. 38950-01J0, 185 cm, J-Tip, Single Pack. Sterile EO. Manufactured for: Boston Scientific Corporation, One Boston Scientific Place, Natick, MN 01760 USA. Boston Scientific Guide Wires are steerable guide wires available in a variety of lengths and diameters. The distal tip is shapeable, or, alternatively, is available in a preshaped ''J'' tip.

Reason

Boston Scientific discovered through an internal inspection process that the PTFE (polytetrafluoroethylene) coating is not consistently adhering to the transition area of the distal end of the guide wires which may cause PTFE to flake off.

Action

On May 30, 2006, Boston Scientific notified customers of the IQ Guide Wire manufactured from February 3, 2006, through March 15, 2006. BSC requested a reply whether or not they still have product in their institution. Replacements will be issued for returned product. UPDATE: A second recall notification was sent to affected customers 07/07/06. This second notification included Inner Pouch Number (UPN) and Inner Pouch Lot/Batch #.

Distribution

Nationwide (AR, AZ, FL, GA, IA, ID, IL, IN, KY, LA, MA, MI, MN, MO, NC, NH, NJ, NM, NV, OH, OK, PA, TN, TX, UT, VA, WI, WV) and Puerto Rico. Chile, Germany, Taiwan.

Quantity

1455 units total