FDA Recall Terminated

Procedure kits include devices from the affected lots distributed OUS only: (1)BOX KIT00529H KIT COLECISTECTOMIA ACCESO (2)BOX KIT00680 KIT COLECISTECTOMIA- (3)BOX KIT00681 KIT APENDICECTOMIA (4)BOX KIT00897 KIT COLON IZQUIERDO (5)BOX LAPL1 ZESTAW DO CHOLECYSTEKTOMII (6)BOX PST01134 LAP CHOLE KIT ST JOHN S (7)KIT00599 LAPARO GENERICO X1 (8)KIT00608 APENDICECTOMIA LAP X1 (9)KIT2176CV BORNHOLM LAP INGUINAL HERNIA (10)KIT2951V CHOLECYSTECT BORNHOLM (11KIT2963 APPENDECTOMY KIT X1 (12)KITM025 M'LANDS MR NASSAR LAP CHOLEX1 The Covidien Endo Grasp" Auto Suture" grasper 5 mm instrument has applications in a variety of gynecologic, general, urologic, thoracic and endoscopic procedures for temporary grasping/clamping of tissue and small tubular structures. 1)BOX KIT00529H KIT COLECISTECTOMIA ACCESO Lot Number: 0217118756 (2)BOX KIT00680 KIT COLECISTECTOMIA- 0217117865 0217118746 0217117863 0217117858 0217117864 0217117862 0217117861 0217117859 0217117857 0217117860 (3)BOX KIT00681 KIT APENDICECTOMIA 0217118750 0217118749 0217118748 0217118747 (4)BOX KIT00897 KIT COLON IZQUIERDO 0217134109 (5)BOX LAPL1 ZESTAW DO CHOLECYSTEKTOMII 0217118761 (6)BOX PST01134 LAP CHOLE KIT ST JOHN S 0217119261, 0217119259 (7)KIT00599 LAPARO GENERICO X1 0217130753, 0217130754 (8)KIT00608 APENDICECTOMIA LAP X1 0217132887 (9)KIT2176CV BORNHOLM LAP INGUINAL HERNIA 0217123007 (10)KIT2951V CHOLECYSTECT BORNHOLM 0217123010 (11KIT2963 APPENDECTOMY KIT X1 0217123009 (12)KITM025 M'LANDS MR NASSAR LAP CHOLEX1 0217127223 (1)BOX KIT00529H KIT COLECISTECTOMIA ACCESO Lot Number: 0217118756 (2)BOX KIT00680 KIT COLECISTECTOMIA- 0217117865 0217118746 0217117863 0217117858 0217117864 0217117862 0217117861 0217117859 0217117857 0217117860 (3)BOX KIT00681 KIT APENDICECTOMIA 0217118750 0217118749 0217118748 0217118747 (4)BOX KIT00897 KIT COLON IZQUIERDO 0217134109 (5)BOX LAPL1 ZESTAW DO CHOLECYSTEKTOMII 0217118761 (6)BOX PST01134 LAP CHOLE KIT ST JOHN S 0217119261, 0217119259 (7)KIT00599 LAPARO GENERICO X1 0217130753, 0217130754 (8)KIT00608 APENDICECTOMIA LAP X1 0217132887 (9)KIT2176CV BORNHOLM LAP INGUINAL HERNIA 0217123007 (10)KIT2951V CHOLECYSTECT BORNHOLM 0217123010 (11KIT2963 APPENDECTOMY KIT X1 0217123009 (12)KITM025 M'LANDS MR NASSAR LAP CHOLEX1 0217127223 (1)BOX KIT00529H KIT COLECISTECTOMIA ACCESO Lot Number: 0217118756 (2)BOX KIT00680 KIT COLECISTECTOMIA- 0217117865 0217118746 0217117863 0217117858 0217117864 0217117862 0217117861 0217117859 0217117857 0217117860 (3)BOX KIT00681 KIT APENDICECTOMIA 0217118750 0217118749 0217118748 0217118747 (4)BOX KIT00897 KIT COLON IZQUIERDO 0217134109 (5)BOX LAPL1 ZESTAW DO CHOLECYSTEKTOMII 0217118761 (6)BOX PST01134 LAP CHOLE KIT ST JOHN S 0217119261, 0217119259 (7)KIT00599 LAPARO GENERICO X1 0217130753, 0217130754 (8)KIT00608 APENDICECTOMIA LAP X1 0217132887 (9)KIT2176CV BORNHOLM LAP INGUINAL HERNIA 0217123007 (10)KIT2951V CHOLECYSTECT BORNHOLM 0217123010 (11KIT2963 APPENDECTOMY KIT X1 0217123009 (12)KITM025 M'LANDS MR NASSAR LAP CHOLEX1 0217127223

Recall: Z-1213-2019 · Initiated March 13, 2019

Recall

Recall Number
Z-1213-2019
Event Number
82426
Firm
COVIDIEN LLC
FEI Number
1282497
Product Code
HDA
Status
Terminated
Root Cause
Process control
Initiated
March 13, 2019
Terminated
April 23, 2020
Address
15 Hampshire St, Mansfield, MA, 02048-1113

Description

Procedure kits include devices from the affected lots distributed OUS only: (1)BOX KIT00529H KIT COLECISTECTOMIA ACCESO (2)BOX KIT00680 KIT COLECISTECTOMIA- (3)BOX KIT00681 KIT APENDICECTOMIA (4)BOX KIT00897 KIT COLON IZQUIERDO (5)BOX LAPL1 ZESTAW DO CHOLECYSTEKTOMII (6)BOX PST01134 LAP CHOLE KIT ST JOHN S (7)KIT00599 LAPARO GENERICO X1 (8)KIT00608 APENDICECTOMIA LAP X1 (9)KIT2176CV BORNHOLM LAP INGUINAL HERNIA (10)KIT2951V CHOLECYSTECT BORNHOLM (11KIT2963 APPENDECTOMY KIT X1 (12)KITM025 M'LANDS MR NASSAR LAP CHOLEX1 The Covidien Endo Grasp" Auto Suture" grasper 5 mm instrument has applications in a variety of gynecologic, general, urologic, thoracic and endoscopic procedures for temporary grasping/clamping of tissue and small tubular structures. 1)BOX KIT00529H KIT COLECISTECTOMIA ACCESO Lot Number: 0217118756 (2)BOX KIT00680 KIT COLECISTECTOMIA- 0217117865 0217118746 0217117863 0217117858 0217117864 0217117862 0217117861 0217117859 0217117857 0217117860 (3)BOX KIT00681 KIT APENDICECTOMIA 0217118750 0217118749 0217118748 0217118747 (4)BOX KIT00897 KIT COLON IZQUIERDO 0217134109 (5)BOX LAPL1 ZESTAW DO CHOLECYSTEKTOMII 0217118761 (6)BOX PST01134 LAP CHOLE KIT ST JOHN S 0217119261, 0217119259 (7)KIT00599 LAPARO GENERICO X1 0217130753, 0217130754 (8)KIT00608 APENDICECTOMIA LAP X1 0217132887 (9)KIT2176CV BORNHOLM LAP INGUINAL HERNIA 0217123007 (10)KIT2951V CHOLECYSTECT BORNHOLM 0217123010 (11KIT2963 APPENDECTOMY KIT X1 0217123009 (12)KITM025 M'LANDS MR NASSAR LAP CHOLEX1 0217127223 (1)BOX KIT00529H KIT COLECISTECTOMIA ACCESO Lot Number: 0217118756 (2)BOX KIT00680 KIT COLECISTECTOMIA- 0217117865 0217118746 0217117863 0217117858 0217117864 0217117862 0217117861 0217117859 0217117857 0217117860 (3)BOX KIT00681 KIT APENDICECTOMIA 0217118750 0217118749 0217118748 0217118747 (4)BOX KIT00897 KIT COLON IZQUIERDO 0217134109 (5)BOX LAPL1 ZESTAW DO CHOLECYSTEKTOMII 0217118761 (6)BOX PST01134 LAP CHOLE KIT ST JOHN S 0217119261, 0217119259 (7)KIT00599 LAPARO GENERICO X1 0217130753, 0217130754 (8)KIT00608 APENDICECTOMIA LAP X1 0217132887 (9)KIT2176CV BORNHOLM LAP INGUINAL HERNIA 0217123007 (10)KIT2951V CHOLECYSTECT BORNHOLM 0217123010 (11KIT2963 APPENDECTOMY KIT X1 0217123009 (12)KITM025 M'LANDS MR NASSAR LAP CHOLEX1 0217127223 (1)BOX KIT00529H KIT COLECISTECTOMIA ACCESO Lot Number: 0217118756 (2)BOX KIT00680 KIT COLECISTECTOMIA- 0217117865 0217118746 0217117863 0217117858 0217117864 0217117862 0217117861 0217117859 0217117857 0217117860 (3)BOX KIT00681 KIT APENDICECTOMIA 0217118750 0217118749 0217118748 0217118747 (4)BOX KIT00897 KIT COLON IZQUIERDO 0217134109 (5)BOX LAPL1 ZESTAW DO CHOLECYSTEKTOMII 0217118761 (6)BOX PST01134 LAP CHOLE KIT ST JOHN S 0217119261, 0217119259 (7)KIT00599 LAPARO GENERICO X1 0217130753, 0217130754 (8)KIT00608 APENDICECTOMIA LAP X1 0217132887 (9)KIT2176CV BORNHOLM LAP INGUINAL HERNIA 0217123007 (10)KIT2951V CHOLECYSTECT BORNHOLM 0217123010 (11KIT2963 APPENDECTOMY KIT X1 0217123009 (12)KITM025 M'LANDS MR NASSAR LAP CHOLEX1 0217127223

Reason

Sterilization method used was not consistent with the labeling and the approved sterilization method

Action

Medtronic notified consignees via an "URGENT MEDICAL DEVICE RECALL" letter by Federal Express and/or certified mail on March 13, 2019. The letter informs customers of the recall of twenty-one production lot of Covidien Endo Clinch" and Endo Grasp" Auto Suture" graspers 5mm as the sterilization method used for these lots was not consistent with the labeling and the approved sterilization method. Customers are instructed to take the following actions: 1. Please immediately quarantine and discontinue use of the affected item codes and lots listed above. 2. Please return affected product as indicated below. 3. If you have distributed the Endo Clinch" II or Endo Grasp" Auto Suture" graspers 5mm listed above, please promptly forward the information from this letter to those recipients. 4. Complete the Recalled Product Return Form even if you do not have inventory. -----If you have inventory email to: [email protected] fax to: 800-895-6140. -----Exception: Customers with zero inventory, fax to 203-492-7719 or email to [email protected] If purchased by distributor, fax to 203-492-7719 or email [email protected]. If you have any questions or concerns, please do not hesitate to contact your Medtronic representative or Customer Service at 800-962-9888, option 2.

Distribution

Worldwide Distribution: US (Nationwide) and to countries of: Canada, Chile, Costa Rica, Guatemala, Japan, Mexico, Panama, Slovakia, Spain, Switzerland and Turkey.

Quantity

148 units