FDA Recall Terminated

Mar Cor Purification Millenium HX Portable Reverse Osmosis System. Model 3025975. Product Usage: Water Purification System is intended to be used as a dialysis accessory to produce water through reverse osmosis for use with hemodialysis equipment..

Recall: Z-1213-2014 · Initiated June 26, 2013

Recall

Recall Number
Z-1213-2014
Event Number
67517
Firm
Mar Cor Purification
FEI Number
3006475063
Product Code
FIP
Status
Terminated
Root Cause
Component design/selection
Initiated
June 26, 2013
Posted
March 14, 2014
Terminated
January 22, 2015
Address
14550 28th Ave N, Minneapolis, MN, 55447-4817

Description

Mar Cor Purification Millenium HX Portable Reverse Osmosis System. Model 3025975. Product Usage: Water Purification System is intended to be used as a dialysis accessory to produce water through reverse osmosis for use with hemodialysis equipment..

Reason

High inlet water pressure beyond specification and the solenoid valve does not meet application which could result in a delay in treatment.

Action

Mar Cor sent an Urgent Medical Device Correction letter dated March 5, 2014 to all affected customers. The letter identifies the affected devices, what the issue is and tells customers to contact Mar Cor to schedule correction. A response form was asked to be completed and returned. For questions contact Scott Nichols, Project Coordinator/Service Trainer at 1-800-633-3080.

Distribution

USA Nationwide Distribution including Puerto Rico.

Quantity

232 units