FDA Recall Terminated

XVIVO Organ Chamber REF 19020

Recall: Z-1212-2022 · Initiated April 8, 2022

Recall

Recall Number
Z-1212-2022
Event Number
90143
Firm
XVIVO PERFUSION AB Massans Gata 10 Goteborg Sweden
FEI Number
3007710603
Product Code
PHO
Status
Terminated
Root Cause
Under Investigation by firm
Initiated
April 8, 2022
Terminated
January 30, 2025

Description

XVIVO Organ Chamber REF 19020

Reason

The weld seal of the primary pouch was incomplete, and the sterility of the product cannot be ensured/guaranteed.

Action

On or about 04/11/2022, XVIVO Inc. issued an "URGENT- Field Safety Notice" for its XVIVO Organ Chamber via email to customers informing them XVIVO has identified that the primary package for a number of batches may have notable defects in the weld of the bag. The notable defects are unsealed sections of the weld (i.e. the bag is not completely closed). Unsealed sections in the weld may compromise the sterility of the product and therefore it is important to NOT use a product showing this defect. Customer are instructed to: 1. Carefully check the primary package for unsealed sections. A follow-up "URGENT:-Medical Device Correction" issued on or about 05/06/2022, this letter provides photos of example of unsealed welds. 2. If unsealed sections are noted, do not use the product and contact XVIVO Customer Support at [email protected] for further action. 3. if not defects are noted, the product may be used as intended. Any questions or additional information, contact Customer Support at [email protected] or telephone number +1-303-395-9171.

Distribution

U.S.: CA, FL, MA, NC, NY, OH, PA, and TX O.U.S.: Austria, Canada, France, Ireland, Italy, and the Netherland

Quantity

80 devices