FDA Recall Open, Classified

Sprint 200 wheeled stretcher, Item numbers 1ES200-7, 1ES210-1, and 1ES211-1.

Recall: Z-1211-2023 · Initiated January 18, 2023

Recall

Recall Number
Z-1211-2023
Event Number
91492
Firm
Linet Spol. S.r.o. Zelevcice 5 Zelevcice Czech Republic
FEI Number
3008058231
Product Code
FPO
Status
Open, Classified
Root Cause
Labeling design
Initiated
January 18, 2023
Posted
March 4, 2023

Description

Sprint 200 wheeled stretcher, Item numbers 1ES200-7, 1ES210-1, and 1ES211-1.

Reason

Labeling being updated to assure user is securing the siderail to a locked position to prevent potential patient fall and injury.

Action

The firm issued "URGENT MEDICAL DEVICE CORRECTION" recall letters dated 1/17/2023 via email on 1/18/2023. The letter explains the issue that an audible clicking sound can be heard when the side rail is raised, but this sound does not indicate the side rail has locked into the raised position. The consignee is to take immediate action to notify caregivers using the Sprint 200 of this potential hazard and proper use for securing a side rail in the raised position. The consignee is informed that Linet will be adding additional safety warning labels located at each side rail latch (2 per side of stretcher), add additional safety warning labels to both side rails (1 per side of stretcher), and that additional training material and an updated copy of the Sprint 200 Emergency Stretcher Instructions for Use and Technical Description manual is available using the QR code in the letter or by following a link to access the materials. The firm's service department will be contacting the consignees to schedule an appointment to perform the corrections. Photos of the additional warning labels are included in the letter. Please contact LINET service department at 877-815-9897 with immediate questions or concerns.

Distribution

US Distribution was made to AL, CA, FL, IA, KY, LA, MN, NV, NC, OK, TX, VA, WA, and WV. There was no government/military distribution. Foreign distribution was made to Australia, Bahrain, Belgium, Germany, Great Britain, Italy, Netherland, Singapore, South Africa, Switzerland, and United Emirates Canada.

Quantity

335 stretchers (310 US, 25 OUS)